Feasibility and performance of spin-echo EPI MR elastography at 3 Tesla for staging hepatic fibrosis in the presence of hepatic iron overload

David Sgier, Daniel Stocker, Christoph Jüngst, Melanie Renzulli, Hanna Biletska-Hanchorova, Achim Weber, Stephan Kannengiesser, Christoph Gubler, Caecilia S Reiner iD

Abdominal Radiology · 2024

DOI: 10.1007/s00261-023-04160-0 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Abdominal Radiology, 2024. Verfasst von David Sgier und 8 weiteren Autor:innen. Beteiligte Einrichtung: Institute of Diagnostic and Interventional Radiology, University Hospital Zurich, Raemistrasse 100, 8091, Zurich, Switzerland. Hauptthemen laut MEDLINE: Elasticity Imaging Techniques, Liver Cirrhosis, Feasibility Studies, Iron Overload. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: University of Zurich.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To assess the feasibility and performance of MR elastography (MRE) for quantifying liver fibrosis in patients with and without hepatic iron overload.

Methods

This retrospective single-center study analyzed 139 patients who underwent liver MRI at 3 Tesla including MRE (2D spin-echo EPI sequence) and R2* mapping for liver iron content (LIC) estimation. MRE feasibility and diagnostic performance between patients with normal and elevated LIC were compared.

Results

Patients with elevated LIC (21%) had significantly higher MRE failure rates (24.1% vs. 3.6%, p < 0.001) compared to patients with normal LIC (79%). For those with only insignificant to mild iron overload (LIC < 5.4 mg/g; 17%), MRE failure rate did not differ significantly from patients without iron overload (8.3% vs. 3.6%, p = 0.315). R2* predicted MRE failure with fair accuracy at a threshold of R2* ≥ 269 s −1 (LIC of approximately 4.6 mg/g). MRE showed good diagnostic performance for detecting significant (≥ F2) and severe fibrosis (≥ F3) in patients without (AUC 0.835 and 0.900) and with iron overload (AUC 0.818 and 0.889) without significant difference between the cohorts ( p = 0.884 and p = 0.913). For detecting cirrhosis MRE showed an excellent diagnostic performance in both groups (AUC 0.944 and 1.000, p = 0.009).

Conclusion

Spin-echo EPI MRE at 3 Tesla is feasible in patients with mild iron overload with good to excellent performance for detecting hepatic fibrosis with a failure rate comparable to patients without iron overload.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Elasticity Imaging TechniquesLiver CirrhosisFeasibility StudiesIron OverloadHumansMaleFemaleMiddle AgedRetrospective StudiesAdultAgedEcho-Planar ImagingLiverMagnetic Resonance Imaging

Stichwörter der Autor:innen

Elasticity imaging techniquesIron overloadLiver cirrhosisSpin-echo EPI MR elastography

Autor:innen und Institutionen

  • Institute of Diagnostic and Interventional Radiology, University Hospital Zurich, Raemistrasse 100, 8091, Zurich, Switzerland.
  • University of Zurich, Zurich, Switzerland.
  • Department of Gastroenterology and Hepatology, University Hospital Zurich, Zurich, Switzerland.
  • Institute of Pathology and Molecular Pathology, University Hospital Zurich, Zurich, Switzerland.
  • MR Application Predevelopment, Siemens Healthcare GmbH, Erlangen, Germany.
  • Department of Gastroenterology and Hepatology, Stadtspital Triemli, Zurich, Switzerland.

ORCID-Profile: Caecilia S Reiner

Förderung

  • University of Zurich

Zitieren

APA

Sgier, D., Stocker, D., Jüngst, C., Renzulli, M., Biletska-Hanchorova, H., Weber, A., Kannengiesser, S., Gubler, C., & Reiner, C. S. (2024). Feasibility and performance of spin-echo EPI MR elastography at 3 Tesla for staging hepatic fibrosis in the presence of hepatic iron overload. Abdominal Radiology. https://doi.org/10.1007/s00261-023-04160-0

Vancouver

Sgier D, Stocker D, Jüngst C, Renzulli M, Biletska-Hanchorova H, Weber A, et al. Feasibility and performance of spin-echo EPI MR elastography at 3 Tesla for staging hepatic fibrosis in the presence of hepatic iron overload. Abdominal Radiology. 2024. doi:10.1007/s00261-023-04160-0

BibTeX
@article{Sgier2024,
  author = {Sgier, David and Stocker, Daniel and Jüngst, Christoph and Renzulli, Melanie and Biletska-Hanchorova, Hanna and Weber, Achim and Kannengiesser, Stephan and Gubler, Christoph and Reiner, Caecilia S},
  title = {Feasibility and performance of spin-echo EPI MR elastography at 3 Tesla for staging hepatic fibrosis in the presence of hepatic iron overload},
  journal = {Abdominal Radiology},
  year = {2024},
  doi = {10.1007/s00261-023-04160-0}
}
Kostenlos starten →