Steerable DROP-IN radioguidance during minimal-invasive non-robotic cervical and endometrial sentinel lymph node surgery
European Journal of Nuclear Medicine and Molecular Imaging · 2024
DOI: 10.1007/s00259-023-06589-3 ↗Kurz gefasst
Klinische Studie (Phase II) in der Zeitschrift European Journal of Nuclear Medicine and Molecular Imaging, 2024. Verfasst von Matthias N van Oosterom und 9 weiteren Autor:innen. Beteiligte Einrichtung: Interventional Molecular Imaging Laboratory, Department of Radiology, Leiden University Medical Center, Leiden, The Netherlands. m.n.van_oosterom@lum…. Hauptthemen laut MEDLINE: Endometrial Neoplasms, Sentinel Lymph Node, Uterine Cervical Neoplasms. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 43 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Stichting voor de Technische Wetenschappen, Sociedad Española de Medicina Nuclear e Imagen Molecular.
Abstract (Originalsprache)
Purpose
The recently introduced tethered DROP-IN gamma probe has revolutionized the way robotic radioguided surgery is performed, fully exploiting the nature of steerable robotic instruments. Given this success, the current first-in-human study investigates if the DROP-IN can also provide benefit in combination with steerable non-robotic instruments during conventional laparoscopic surgery, showing equivalence or even benefit over a traditional rigid gamma probe.
Methods
The evaluation was performed in ten patients during laparoscopic cervical ( n = 4) and endometrial ( n = 6) cancer sentinel lymph node (SLN) procedures. Surgical guidance was provided using the hybrid, or bi-modal, SLN tracer ICG- 99m Tc-nanocolloid. SLN detection was compared between the traditional rigid laparoscopic gamma probe, the combination of a DROP-IN gamma probe and a steerable laparoscopic instrument (LaproFlex), and fluorescence imaging.
Results
The gynecologists experienced an enlarged freedom of movement when using the DROP-IN + LaproFlex combination compared to the rigid laparoscopic probe, making it possible to better isolate the SLN signal from background signals. This did not translate into a change in the SLN find rate yet. In both cervical and endometrial cancer combined, the rigid probe and DROP-IN + LaproFlex combination provided an equivalent detection rate of 96%, while fluorescence provided 85%.
Conclusion
We have successfully demonstrated the in-human use of steerable DROP-IN radioguidance during laparoscopic cervical and endometrial cancer SLN procedures, expanding the utility beyond robotic procedures. Indicating an improved surgical experience, these findings encourage further investigation and consideration on a path towards routine clinical practice and improved patient outcome.
Trial registration
HCB/2021/0777 and NCT04492995; https://clinicaltrials.gov/study/NCT04492995
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Interventional Molecular Imaging Laboratory, Department of Radiology, Leiden University Medical Center, Leiden, The Netherlands. m.n.van_oosterom@lumc.nl.
- Gynecology Oncology Unit, Institute Clínic of Gynecology, Obstetrics, and Neonatology, Hospital Clinic of Barcelona, Barcelona, Spain.
- Fundació de Recerca Clínic Barcelona - Institut d'Investigacions Biomèdiques August Pi I Sunyer (FRCB-IDIBAPS), Barcelona, Spain.
- Faculty of Medicine, University of Barcelona, Barcelona, Spain.
- Nuclear Medicine Department, Hospital Clínic of Barcelona, Barcelona, Spain.
- Nuclear Medicine Department, Hospital Universitario Puerta del Mar, Cádiz, Spain.
ORCID-Profile: Matthias N van Oosterom, Berta Diaz-Feijóo, Andrés Perissinotti, Ariel Glickman, Tiermes Marina, Aureli Torné, Fijs W B van Leeuwen, Sergi Vidal-Sicart
Förderung
- Stichting voor de Technische Wetenschappen
- Sociedad Española de Medicina Nuclear e Imagen Molecular
Zitieren
APA
van Oosterom, M. N., Diaz-Feijóo, B., Santisteban, M. I., Sánchez-Izquierdo, N., Perissinotti, A., Glickman, A., Marina, T., Torné, A., van Leeuwen, F. W. B., & Vidal-Sicart, S. (2024). Steerable DROP-IN radioguidance during minimal-invasive non-robotic cervical and endometrial sentinel lymph node surgery. European Journal of Nuclear Medicine and Molecular Imaging. https://doi.org/10.1007/s00259-023-06589-3
Vancouver
van Oosterom MN, Diaz-Feijóo B, Santisteban MI, Sánchez-Izquierdo N, Perissinotti A, Glickman A, et al. Steerable DROP-IN radioguidance during minimal-invasive non-robotic cervical and endometrial sentinel lymph node surgery. European Journal of Nuclear Medicine and Molecular Imaging. 2024. doi:10.1007/s00259-023-06589-3
BibTeX
@article{vanOosterom2024,
author = {van Oosterom, Matthias N and Diaz-Feijóo, Berta and Santisteban, Maria Isabel and Sánchez-Izquierdo, Núria and Perissinotti, Andrés and Glickman, Ariel and Marina, Tiermes and Torné, Aureli and van Leeuwen, Fijs W B and Vidal-Sicart, Sergi},
title = {Steerable DROP-IN radioguidance during minimal-invasive non-robotic cervical and endometrial sentinel lymph node surgery},
journal = {European Journal of Nuclear Medicine and Molecular Imaging},
year = {2024},
doi = {10.1007/s00259-023-06589-3}
}
EUVIMED