Mitigating SUV uncertainties using total body PET imaging
European Journal of Nuclear Medicine and Molecular Imaging · 2024
DOI: 10.1007/s00259-023-06503-x ↗Kurz gefasst
Fachartikel in der Zeitschrift European Journal of Nuclear Medicine and Molecular Imaging, 2024. Verfasst von Charlotte L C Smith und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology and Nuclear Medicine, Amsterdam UMC location Vrije Universiteit Amsterdam, De Boelelaan 1117 1081 HV, Amsterdam, The Netherla…. Hauptthemen laut MEDLINE: Fluorodeoxyglucose F18, Positron Emission Tomography Computed Tomography. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Siemens Healthineers.
Abstract (Originalsprache)
Purpose
Standardised uptake values (SUV) are commonly used to quantify 18 F-FDG lesion uptake. However, SUVs may suffer from several uncertainties and errors. Long-axial field-of-view (LAFOV) PET/CT systems might enable image-based quality control (QC) by deriving 18 F-FDG activity and weight from total body (TB) 18 F-FDG PET images. In this study, we aimed to develop these image-based QC to reduce errors and mitigate SUV uncertainties.
Methods
Twenty-five out of 81 patient scans from a LAFOV PET/CT system were used to determine regression fits for deriving of image-derived activity and weight. Thereafter, the regression fits were applied to 56 independent 18 F-FDG PET scans from the same scanner to determine if injected activity and weight could be obtained accurately from TB and half-body (HB) scans. Additionally, we studied the impact of image-based values on the precision of liver SUVmean and lesion SUVpeak. Finally, 20 scans were acquired from a short-axial field-of-view (SAFOV) PET/CT system to determine if the regression fits also applied to HB scans from a SAFOV system.
Results
Both TB and HB 18 F-FDG activity and weight significantly predicted reported injected activity ( r = 0.999; r = 0.984) and weight ( r = 0.999; r = 0.987), respectively. After applying the regression fits, 18 F-FDG activity and weight were accurately derived within 4.8% and 3.2% from TB scans and within 4.9% and 3.1% from HB, respectively. Image-derived values also mitigated liver and lesion SUV variability compared with reported values. Moreover, 18 F-FDG activity and weight obtained from a SAFOV scanner were derived within 6.7% and 4.5%, respectively.
Conclusion
18 F-FDG activity and weight can be derived accurately from TB and HB scans, and image-derived values improved SUV precision and corrected for lesion SUV errors. Therefore, image-derived values should be included as QC to generate a more reliable and reproducible quantitative uptake measurement.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Radiology and Nuclear Medicine, Amsterdam UMC location Vrije Universiteit Amsterdam, De Boelelaan 1117 1081 HV, Amsterdam, The Netherlands. c.l.c.smith@amsterdamumc.nl.
- Cancer Center Amsterdam, Imaging and Biomarkers, Amsterdam, The Netherlands. c.l.c.smith@amsterdamumc.nl.
- Department of Radiology and Nuclear Medicine, Amsterdam UMC location Vrije Universiteit Amsterdam, De Boelelaan 1117 1081 HV, Amsterdam, The Netherlands.
- Cancer Center Amsterdam, Imaging and Biomarkers, Amsterdam, The Netherlands.
ORCID-Profile: Charlotte L C Smith
Förderung
- Siemens Healthineers
Zitieren
APA
Smith, C. L. C., Zwezerijnen, G. J. C., den Hollander, M. E., Weijland, J., Yaqub, M., & Boellaard, R. (2024). Mitigating SUV uncertainties using total body PET imaging. European Journal of Nuclear Medicine and Molecular Imaging. https://doi.org/10.1007/s00259-023-06503-x
Vancouver
Smith CLC, Zwezerijnen GJC, den Hollander ME, Weijland J, Yaqub M, Boellaard R. Mitigating SUV uncertainties using total body PET imaging. European Journal of Nuclear Medicine and Molecular Imaging. 2024. doi:10.1007/s00259-023-06503-x
BibTeX
@article{Smith2024,
author = {Smith, Charlotte L C and Zwezerijnen, Gerben J C and den Hollander, Marijke E and Weijland, Jolijn and Yaqub, Maqsood and Boellaard, Ronald},
title = {Mitigating SUV uncertainties using total body PET imaging},
journal = {European Journal of Nuclear Medicine and Molecular Imaging},
year = {2024},
doi = {10.1007/s00259-023-06503-x}
}
EUVIMED