Reducing Metal Artifacts in Clinical Photon Counting Detector Computed Tomography—A Phantom Study of an Exemplary Total Hip Arthroplasty
Skeletal Radiology · 2025
DOI: 10.1007/s00256-024-04820-2 ↗Kurz gefasst
Fachartikel in der Zeitschrift Skeletal Radiology, 2025. Verfasst von Roy P Marcus und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Balgrist University Hospital, Zurich, Switzerland. Roy.marcus@balgrist.ch. Hauptthemen laut MEDLINE: Artifacts, Hip Prosthesis, Tomography, X-Ray Computed, Arthroplasty, Replacement, Hip. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 29 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: University of Zurich.
Abstract (Originalsprache)
Objective
To examine how different photon-counting detector (PCD) CT scanning and reconstruction methods affect the volume of metal artifacts and image quality for a hip prosthesis phantom.
Methods
A titanium and cobalt-chromium-molybdenum-alloy total hip prosthesis phantom was scanned using a clinical PCD-CT with a constant tube potential (140 kV) and Computed-Tomography-Dose- Index (7 mGy). Different scan settings were used: with/without tin-filter (Sn), with/without ultra-high resolution (UHR), both individually and combined, resulting in four modes: Quantumplus (Standard), UHR Quantumplus (HighRes), QuantumSn (Standard-Tin) and UHR QuantumSn (HighRes-Tin). Reconstructions included virtual monoenergetic images (VMI) spanning 40–190 keV and polychromatic images, with/without iterative metal artifact reduction (MAR). Artifact volumes rendered in a 3D-printing software were quantified in milliliters (ml), and image quality was evaluated using a Likert score.
Results
Polychromatic reconstruction: Tin-filter reduced artifact volumes (298 (Standard-Tin) vs. 347 ml (Standard) and 310 (HighRes-Tin) vs. 360 ml (HighRes)). The smallest artifact volume was measured in HighRes MAR (150 ml). VMI reconstruction: The smallest artifact volume was measured in Standard 130 keV (150 ml) and HighRes 130 keV (164 ml) and in Standard-Tin 120 keV (169 ml) and HighRes-Tin 120 keV (172 ml). MAR further reduced the artifact volumes to 130 ml (Standard 150 keV MAR) and 140 ml (HighRes 160 keV MAR). Image quality was rated best for Standard 65 keV MAR, polychromatic HighRes MAR, Standard 100 keV MAR, polychromatic Standard-tin MAR, HighRes-tin 100 keV and polychromatic HighRes-tin.
Conclusion
Combining tin-filter, UHR and MAR in VMI or polychromatic images achieve the strongest artifact reduction.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Radiology, Balgrist University Hospital, Zurich, Switzerland. Roy.marcus@balgrist.ch.
- Faculty of Medicine, University of Zurich, Zurich, Switzerland. Roy.marcus@balgrist.ch.
- Department of Radiology, Balgrist University Hospital, Zurich, Switzerland.
- Faculty of Medicine, University of Zurich, Zurich, Switzerland.
- Swiss Center for Musculoskeletal Imaging, Balgrist Campus, Zurich, Switzerland.
- Siemens Healthineers, Forchheim, Germany.
Förderung
- University of Zurich
Zitieren
APA
Marcus, R. P., Feuerriegel, G. C., Marth, A. A., Goller, S. S., Nanz, D., Anhaus, J., & Sutter, R. (2025). Reducing Metal Artifacts in Clinical Photon Counting Detector Computed Tomography—A Phantom Study of an Exemplary Total Hip Arthroplasty. Skeletal Radiology. https://doi.org/10.1007/s00256-024-04820-2
Vancouver
Marcus RP, Feuerriegel GC, Marth AA, Goller SS, Nanz D, Anhaus J, et al. Reducing Metal Artifacts in Clinical Photon Counting Detector Computed Tomography—A Phantom Study of an Exemplary Total Hip Arthroplasty. Skeletal Radiology. 2025. doi:10.1007/s00256-024-04820-2
BibTeX
@article{Marcus2025,
author = {Marcus, Roy P and Feuerriegel, Georg C and Marth, Adrian A and Goller, Sophia S and Nanz, Daniel and Anhaus, Julian and Sutter, Reto},
title = {Reducing Metal Artifacts in Clinical Photon Counting Detector Computed Tomography—A Phantom Study of an Exemplary Total Hip Arthroplasty},
journal = {Skeletal Radiology},
year = {2025},
doi = {10.1007/s00256-024-04820-2}
}
EUVIMED