EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Radiographic calibration for pubic diastasis assessment in bladder exstrophy-epispadias complex: a phantom study

S J Back, D A Weiss, B Marshall, E Akbari, M Mackey, E Hinton, B D Horn, M Kidd, M L Francavilla

Pediatric Radiology · 2024

DOI: 10.1007/s00247-024-05972-y ↗

Kurz gefasst

Fachartikel in der Zeitschrift Pediatric Radiology, 2024. Verfasst von S J Back und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Children's Hospital of Philadelphia, Philadelphia, PA, USA. backs@chop.edu. Hauptthemen laut MEDLINE: Bladder Exstrophy, Phantoms, Imaging, Epispadias, Tomography, X-Ray Computed. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 19 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

The assessment of pubic diastasis is important for the surgical planning of patients with bladder exstrophy-epispadias complex. Understanding how the diastasis changes during surgical follow-up may help predict patient morbidity. Radiography can follow diastasis but may be affected by patient and technical imaging factors including body size, imaging protocol, and equipment. Using imaging calibration and anatomic ratios may mitigate differences due to these aspects.

Objective

Use imaging phantoms to assess the effect of radiographic calibration on measurements of pubic diastasis and an internal anatomic ratio as a child grows. Materials and methods Radiographic images were obtained of three different sizes of computed tomography phantoms (older child, child, and infant) using three imaging techniques that include the osseous pelvis in children. All phantoms were imaged with abdomen and pelvis techniques. The infant phantom was additionally imaged using a thoracoabdominal technique. These exposures were all repeated with systems from three manufacturers. Linear measurements were made between radiographic markers placed to simulate pubic diastasis and sacral width. A ratio was also created between these distances. Measurements with and without image calibration were made by two pediatric radiologists using rulers placed at the time of image acquisition.

Results

There was excellent interrater agreement for measurements, ICC >0.99. Anterior distances were more affected by magnification than posterior ones with a significant difference between uncalibrated versus calibrated anterior distances ( p =0.04) and not for posterior ones ( p =0.65). There was no difference between radiographic equipment manufacturers without or with calibration ( p values 0.66 to 0.99). There was a significant difference in simulated pubic distance between thoracoabdominal and abdomen ( p =0.04) as well as pelvic ( p =0.04) techniques which resolved with calibration, each p =0.6. The ratio between the simulated pubic diastasis and sacral width differed by phantom size (all p

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Bladder ExstrophyPhantoms, ImagingEpispadiasTomography, X-Ray ComputedHumansCalibrationInfantPubic Symphysis DiastasisFemaleChildMaleReproducibility of ResultsChild, Preschool

Stichwörter der Autor:innen

Bladder exstrophyCalibrationPubic symphysis diastasisRadiography

Autor:innen und Institutionen

  • Department of Radiology, Children's Hospital of Philadelphia, Philadelphia, PA, USA. backs@chop.edu.
  • Perelman School of Medicine, University of Pennsylvania, Philadelphia, PA, USA. backs@chop.edu.
  • Division of Urology, Children's Hospital of Philadelphia, Philadelphia, PA, USA.
  • Perelman School of Medicine, University of Pennsylvania, Philadelphia, PA, USA.
  • Department of Radiology, Children's Hospital of Philadelphia, Philadelphia, PA, USA.
  • Division of Orthopedic Surgery, Children's Hospital of Philadelphia, Philadelphia, PA, USA.

Zitieren

APA

Back, S. J., Weiss, D. A., Marshall, B., Akbari, E., Mackey, M., Hinton, E., Horn, B. D., Kidd, M., & Francavilla, M. L. (2024). Radiographic calibration for pubic diastasis assessment in bladder exstrophy-epispadias complex: a phantom study. Pediatric Radiology. https://doi.org/10.1007/s00247-024-05972-y

Vancouver

Back SJ, Weiss DA, Marshall B, Akbari E, Mackey M, Hinton E, et al. Radiographic calibration for pubic diastasis assessment in bladder exstrophy-epispadias complex: a phantom study. Pediatric Radiology. 2024. doi:10.1007/s00247-024-05972-y

BibTeX
@article{Back2024,
  author = {Back, S J and Weiss, D A and Marshall, B and Akbari, E and Mackey, M and Hinton, E and Horn, B D and Kidd, M and Francavilla, M L},
  title = {Radiographic calibration for pubic diastasis assessment in bladder exstrophy-epispadias complex: a phantom study},
  journal = {Pediatric Radiology},
  year = {2024},
  doi = {10.1007/s00247-024-05972-y}
}
Kostenlos starten →