EUVIMEDEUVIMED entdecken
VergleichsstudieOpen Access · CC BY 4.0

Comparison of actual and automated CT measurements of urinary stone size: a phantom study

Myeong Seong Yoon, Dong-Hyun Jang, Juncheol Lee, Jaehoon Jeong, Do Gwon Kim, Hyojin Lim, Dong Keon Lee, Jaehoon Oh

Urolithiasis · 11.04.2025 · 53 (1)

DOI: 10.1007/s00240-025-01708-1 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Urolithiasis, online erschienen am 11.04.2025. Verfasst von Myeong Seong Yoon und 7 weiteren Autor:innen. Beteiligte Einrichtung: Medical Imaging Processing Lab (MIP), Seoul, Republic of Korea. Hauptthemen laut MEDLINE: Phantoms, Imaging, Tomography, X-Ray Computed, Urinary Calculi. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 23 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Seoul National University.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Urinary stone size is key in determining treatment. Although computed tomography (CT) scans are widely used for diagnosing urinary stones, measurements of stone size obtained from CT images may be inaccurate compared to actual size. Twenty-four urinary stone phantoms were 3D printed at three densities (100, 1000, and 3000 Hounsfield units [HU]) and eight sizes. CT images of the phantoms were taken. Nineteen radiologists and 33 emergency physicians from two institutions measured stone sizes on CT images using mediastinum and bone settings. An automated algorithm segmented regions of interest and estimated stone size using pixel HUs. Mean absolute error (MAE) was assessed for the accuracy of each measurement method against known phantom sizes. For the mediastinum setting, MAEs for 100, 1000, and 3000 HU stone phantoms were 1.05 mm ± 0.06, 1.01 mm ± 0.06, and 2.38 mm ± 0.17, respectively. For the bone setting, MAEs were 0.98 mm ± 0.07, 0.55 mm ± 0.10, and 1.91 mm ± 0.06, respectively. For automated measurements, MAEs were 1.16 mm, 0.21 mm, and 2.10 mm, respectively. Participant-to-participant variability was observed across all measurement settings, regardless of the stone density or window used. For stone size measurements on CT images, the bone setting provided more accurate results than the mediastinum setting. Automated measurement methods, which estimate stone size by outlining its edges, were more accurate than manual measurements for 1000 HU stones, the most common stone density. However, for stones with densities above or below 1000 HU, the accuracy of the automated method may decrease.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Phantoms, ImagingTomography, X-Ray ComputedUrinary CalculiHumansAlgorithms

Stichwörter der Autor:innen

Computed tomographyStone phantomStone sizeUrinary stone

Autor:innen und Institutionen

  • Medical Imaging Processing Lab (MIP), Seoul, Republic of Korea.
  • Seoul National University Bundang Hospital, Seongnam-si, Republic of Korea.
  • Seoul National University College of Medicine, Seoul, Republic of Korea.
  • Hanyang University, Seoul, Republic of Korea.
  • AIDOT Inc, Seoul, Republic of Korea.
  • Seoul National University Bundang Hospital, Seongnam-si, Republic of Korea. goldenegg@snu.ac.kr.

Förderung

  • Seoul National University

Zitieren

APA

Yoon, M. S., Jang, D. H., Lee, J., Jeong, J., Kim, D. G., Lim, H., Lee, D. K., & Oh, J. (2025). Comparison of actual and automated CT measurements of urinary stone size: a phantom study. Urolithiasis, 53(1). https://doi.org/10.1007/s00240-025-01708-1

Vancouver

Yoon MS, Jang DH, Lee J, Jeong J, Kim DG, Lim H, et al. Comparison of actual and automated CT measurements of urinary stone size: a phantom study. Urolithiasis. 2025;53(1). doi:10.1007/s00240-025-01708-1

BibTeX
@article{Yoon2025,
  author = {Yoon, Myeong Seong and Jang, Dong-Hyun and Lee, Juncheol and Jeong, Jaehoon and Kim, Do Gwon and Lim, Hyojin and Lee, Dong Keon and Oh, Jaehoon},
  title = {Comparison of actual and automated CT measurements of urinary stone size: a phantom study},
  journal = {Urolithiasis},
  year = {2025},
  volume = {53},
  number = {1},
  doi = {10.1007/s00240-025-01708-1}
}
Kostenlos starten →