EUVIMEDEUVIMED entdecken
Crossref

Optimal dual-energy computed tomography scan parameters to detect small-sized urinary stones and their composition

Jin-Woo Jung, Jun-Bong Shin, Hyo-Jun Choi, Seongyong Pak, Hyungjin Yang, Byung Il Yoon

Urolithiasis · 2023

DOI: 10.1007/s00240-023-01419-5 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Urolithiasis, 2023. Verfasst von Jin-Woo Jung und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Urology, International St. Mary's Hospital, Catholic Kwandong University College of Medicine, 22711 Simgokro 100Gil 25 Seo-Gu, Incheon,…. Hauptthemen laut MEDLINE: Urinary Calculi, Body Fluids. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 21 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

To investigate the optimal scanning parameters of dual-energy computed tomography (DECT), which can accurately determine sensitivity (the detectability of urinary stones) and accuracy (the composition matching of urinary stones), and to apply them to clinical trials. Fifteen urinary stones were chemically analyzed, and their chemical compositions were considered a reference standard with which we compared the uric acid (UA) and non-UA compositions determined using DECT. The urinary stones were placed inside a bolus and scanned with a dual-source CT scanner under various selected dual-energy conditions (A to X) using various solid water phantom thicknesses. These datasets were analyzed using the Siemens syngo.via software tool (integrated into the CT system) for matching the sensitivity and accuracy assessments. This study showed that 80% of the highest sensitivity (detection of urinary stones) and 92% of the highest accuracy (composition matching of urinary stones) were achieved under condition A (a collimation beam width setting of 2 × 32 × 0.6 mm, an automatic exposure control setting of 80/sn140 peak kilovoltage, and a slice thickness of 0.5/0.5 mm) (P < 0.05). Application of the DECT energy parameters presented in the study will help identify the sensitivity and accuracy of UA and non-UA stone analysis, even in patients with small-sized urinary stones and in conditions difficult for analysis.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Urinary CalculiBody FluidsHumansTomography, X-Ray ComputedPhantoms, ImagingUric Acid

Stichwörter der Autor:innen

Dual-energy computed tomographyNon-uric acidUric acidUrinary stone

Substanzen

Uric Acid

Autor:innen und Institutionen

  • Department of Urology, International St. Mary's Hospital, Catholic Kwandong University College of Medicine, 22711 Simgokro 100Gil 25 Seo-Gu, Incheon, South Korea.
  • Department of Radiation Oncology, Kangwon National University Hospital, Chuncheon, South Korea.
  • Department of Diagnostic Radiology, International St. Mary's Hospital, Catholic Kwandong, Incheon, South Korea.
  • Siemens Healthineers Ltd, Seoul, South Korea.
  • Department of Physics, Korea University Sejong Campus, Sejong, South Korea.
  • Department of Urology, International St. Mary's Hospital, Catholic Kwandong University College of Medicine, 22711 Simgokro 100Gil 25 Seo-Gu, Incheon, South Korea. yoonbi0948@catholic.ac.kr.

Zitieren

APA

Jung, J. W., Shin, J. B., Choi, H. J., Pak, S., Yang, H., & Yoon, B. I. (2023). Optimal dual-energy computed tomography scan parameters to detect small-sized urinary stones and their composition. Urolithiasis. https://doi.org/10.1007/s00240-023-01419-5

Vancouver

Jung JW, Shin JB, Choi HJ, Pak S, Yang H, Yoon BI. Optimal dual-energy computed tomography scan parameters to detect small-sized urinary stones and their composition. Urolithiasis. 2023. doi:10.1007/s00240-023-01419-5

BibTeX
@article{Jung2023,
  author = {Jung, Jin-Woo and Shin, Jun-Bong and Choi, Hyo-Jun and Pak, Seongyong and Yang, Hyungjin and Yoon, Byung Il},
  title = {Optimal dual-energy computed tomography scan parameters to detect small-sized urinary stones and their composition},
  journal = {Urolithiasis},
  year = {2023},
  doi = {10.1007/s00240-023-01419-5}
}
Kostenlos starten →