Automated microextraction by packed sorbent of endocrine disruptors in wastewater using a high-throughput robotic platform followed by liquid chromatography–tandem mass spectrometry
Analytical and Bioanalytical Chemistry · 2023
DOI: 10.1007/s00216-023-04888-0 ↗Kurz gefasst
Fachartikel in der Zeitschrift Analytical and Bioanalytical Chemistry, 2023. Verfasst von Marcio David Bocelli und 3 weiteren Autor:innen. Beteiligte Einrichtung: São Carlos Institute of Chemistry, University of São Paulo, São Carlos, SP, Brazil. Hauptthemen laut MEDLINE: Tandem Mass Spectrometry, Endocrine Disruptors, Wastewater, Water Pollutants, Chemical. Die Arbeit wurde bisher 11-mal zitiert (Crossref) und nennt 58 Literaturangaben. Gefördert von: Coordenação de Aperfeiçoamento de Pessoal de Nível Superior, Conselho Nacional de Desenvolvimento Científico e Tecnológico, Fundação de Amparo à Pesquisa do Estado de São Paulo.
Abstract (Originalsprache)
An automated microextraction by packed sorbent followed by liquid chromatography-tandem mass spectrometry (MEPS-LC-MS/MS) method was developed for the determination of four endocrine disruptors-parabens, benzophenones, and synthetic phenolic antioxidants-in wastewater samples. The method utilizes a lab-made repackable MEPS device and a multi-syringe robotic platform that provides flexibility to test small quantities (2 mg) of multiple extraction phases and enables high-throughput capabilities for efficient method development. The overall performance of the MEPS procedure, including the investigation of influencing variables and the optimization of operational parameters for the robotic platform, was comprehensively studied through univariate and multivariate experiments. Under optimized conditions, the target analytes were effectively extracted from a small sample volume of 1.5 mL, with competitive detectability and analytical confidence. The limits of detection ranged from 0.15 to 0.30 ng L-1, and the intra-day and inter-day relative standard deviations were between 3 and 21%. The method's applicability was successfully demonstrated by determining methylparaben, propylparaben, butylated hydroxyanisole, and oxybenzone in wastewater samples collected from the São Carlos (SP, Brazil) river. Overall, the developed method proved to be a fast, sensitive, reliable, and environmentally friendly analytical tool for water quality monitoring.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- São Carlos Institute of Chemistry, University of São Paulo, São Carlos, SP, Brazil.
- São Carlos Institute of Chemistry, University of São Paulo, São Carlos, SP, Brazil. alvarojsn@iqsc.usp.br.
Förderung
- Coordenação de Aperfeiçoamento de Pessoal de Nível Superior
- Conselho Nacional de Desenvolvimento Científico e Tecnológico
- Fundação de Amparo à Pesquisa do Estado de São Paulo
Zitieren
APA
Bocelli, M. D., Medina, D. A. V., Lanças, F. M., & Dos Santos-Neto, Á. J. (2023). Automated microextraction by packed sorbent of endocrine disruptors in wastewater using a high-throughput robotic platform followed by liquid chromatography–tandem mass spectrometry. Analytical and Bioanalytical Chemistry. https://doi.org/10.1007/s00216-023-04888-0
Vancouver
Bocelli MD, Medina DAV, Lanças FM, Dos Santos-Neto ÁJ. Automated microextraction by packed sorbent of endocrine disruptors in wastewater using a high-throughput robotic platform followed by liquid chromatography–tandem mass spectrometry. Analytical and Bioanalytical Chemistry. 2023. doi:10.1007/s00216-023-04888-0
BibTeX
@article{Bocelli2023,
author = {Bocelli, Marcio David and Medina, Deyber Arley Vargas and Lanças, Fernando Mauro and Dos Santos-Neto, Álvaro José},
title = {Automated microextraction by packed sorbent of endocrine disruptors in wastewater using a high-throughput robotic platform followed by liquid chromatography–tandem mass spectrometry},
journal = {Analytical and Bioanalytical Chemistry},
year = {2023},
doi = {10.1007/s00216-023-04888-0}
}
EUVIMED