EUVIMEDSkopri EUVIMED
Vergleichsstudie

A clinical evaluation of the accuracy of the Nellcor N-100 and Ohmeda 3700 pulse oximeters

W T Cecil, K J Thorpe, E E Fibuch, G F Tuohy

Journal of Clinical Monitoring · 1988-01 · 4 (1) · 31-36

DOI: 10.1007/bf01618105 ↗

At a glance

Vergleichsstudie in Journal of Clinical Monitoring, 1988-01. Written by W T Cecil and 3 co-authors. Institution: Respiratory Care Services, St. Luke's Hospital of Kansas City, MO 64111. Cited 41 times so far (Crossref); 13 references.

AstrattSubjectsDettalji biblijografiċiCiteRelated studies

Abstract (original language)

Two pulse oximeters, the Nellcor N-100 and the Ohmeda 3700, were compared with arterial blood values and with each other in a clinical evaluation of performance. Three hundred twenty-nine simultaneously sampled blood/oximeter data pairs from use of both makes of pulse oximeters on each of 152 test subjects were included in the comparison analysis for each oximeter. Among the patients, disease type and severity and hospital location varied widely. Basic descriptive statistics and linear regression analysis were employed to facilitate comparison. Both oximeters displayed a statistically significant but clinically insignificant bias when compared with arterial blood oxyhemoglobin: -0.31 (P = 0.023) and 0.59 (P = 0.001) for the Ohmeda 3700 pulse oximeter and the Nellcor N-100 pulse oximeter, respectively. Relative to arterial blood oxyhemoglobin, the 95% tolerance intervals were +4.84 to -5.45 (10.3) for the Ohmeda 3700 and +6.94 to -5.76 (12.7) for the Nellcor N-100. Regressed against [oxyhemoglobin + carboxyhemoglobin + methemoglobin] as x, the Nellcor N-100 read y = 0.85(x) + 12.5, r = 0.83, P less than 0.0001, and the Ohmeda 3700 read y = 1.02(x) - 5.3, r = 0.86, P less than 0.0001.

Subjects

MeSH terms (MEDLINE)

FemaleHumansMaleMiddle AgedMonitoring, PhysiologicOximetryRegression Analysis

Authors and institutions

  • Respiratory Care Services, St. Luke's Hospital of Kansas City, MO 64111.

Cite

APA

Cecil, W. T., Thorpe, K. J., Fibuch, E. E., & Tuohy, G. F. (1988). A clinical evaluation of the accuracy of the Nellcor N-100 and Ohmeda 3700 pulse oximeters. Journal of Clinical Monitoring, 4(1), 31-36. https://doi.org/10.1007/bf01618105

Vancouver

Cecil WT, Thorpe KJ, Fibuch EE, Tuohy GF. A clinical evaluation of the accuracy of the Nellcor N-100 and Ohmeda 3700 pulse oximeters. Journal of Clinical Monitoring. 1988;4(1):31-36. doi:10.1007/bf01618105

BibTeX
@article{Cecil1988,
  author = {Cecil, W T and Thorpe, K J and Fibuch, E E and Tuohy, G F},
  title = {A clinical evaluation of the accuracy of the Nellcor N-100 and Ohmeda 3700 pulse oximeters},
  journal = {Journal of Clinical Monitoring},
  year = {1988},
  volume = {4},
  number = {1},
  pages = {31--36},
  doi = {10.1007/bf01618105}
}