EUVIMEDEUVIMED entdecken
Vergleichsstudie

Presentation and evaluation of a new optical sensor for respiratory rate monitoring

M Vegfors, L G Lindberg, H Pettersson, P A Oberg

International Journal of Clinical Monitoring and Computing · 1994

DOI: 10.1007/bf01132363 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift International Journal of Clinical Monitoring and Computing, 1994. Verfasst von M Vegfors und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Anaesthesiology, Linköping University, Sweden. Die Arbeit wurde bisher 35-mal zitiert (Crossref) und nennt 22 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

A new optical sensor for respiratory rate monitoring was simultaneously compared with an acoustic sensor and a transthoracic impedance plethysmograph during normoventilation in the respiratory rate range of 9-17 breaths per minute. The response characteristics of the optical sensor were then measured during simulation of central apnoea and tachypnoea. Visual observation was chosen as the reference method for monitoring the respiratory rate. The measurements were performed in ten healthy volunteers and the respiratory signals recorded on an analogue tape and strip-chart recorder and analysed off-line. The response characteristics of the fibre optic sensor corresponded well with those of the acoustic sensor and impedance plethysmograph. All three methods responded rapidly to an apnoeic event.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

AdultApneaElectrodiagnosisEvaluation Studies as TopicFiber Optic TechnologyHumansMaleMonitoring, PhysiologicPlethysmography, ImpedanceRespiration

Autor:innen und Institutionen

  • Department of Anaesthesiology, Linköping University, Sweden.

Zitieren

APA

Vegfors, M., Lindberg, L. G., Pettersson, H., & Oberg, P. A. (1994). Presentation and evaluation of a new optical sensor for respiratory rate monitoring. International Journal of Clinical Monitoring and Computing. https://doi.org/10.1007/bf01132363

Vancouver

Vegfors M, Lindberg LG, Pettersson H, Oberg PA. Presentation and evaluation of a new optical sensor for respiratory rate monitoring. International Journal of Clinical Monitoring and Computing. 1994. doi:10.1007/bf01132363

BibTeX
@article{Vegfors1994,
  author = {Vegfors, M and Lindberg, L G and Pettersson, H and Oberg, P A},
  title = {Presentation and evaluation of a new optical sensor for respiratory rate monitoring},
  journal = {International Journal of Clinical Monitoring and Computing},
  year = {1994},
  doi = {10.1007/bf01132363}
}
Kostenlos starten →