Calibration‐Free and Highly Sensitive Potentiometric Enzyme‐Based Biosensors

Mitsuo Asai, Nobuyuki Isshiki, Masafumi Takesue, Kuniharu Takei iD

Small Methods · 2025

DOI: 10.1002/smtd.202401010 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Small Methods, 2025. Verfasst von Mitsuo Asai und 3 weiteren Autor:innen. Beteiligte Einrichtung: Performance Chemicals Research, Kao Corporation, Wakayama, 640-8580, Japan. Hauptthemen laut MEDLINE: Biosensing Techniques. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 25 Literaturangaben. Gefördert von: Japan Society for the Promotion of Science, AIP Network Laboratory.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Enzyme‐based amperometric biosensors have become popular for healthcare applications. However, they have been under constant pressure for technological innovation to improve their sensitivity and usability. An ideal biosensor has high sensitivity and calibration‐free characteristics. This study aims to report enzyme‐based glucose and lactate sensors that utilize a proposed “time‐derivative of potential (dOCP/dt)” method, with a further aim being to prove theoretically and experimentally that dOCP/dt values are proportional to substrate concentration. High sensitivity is obtained regardless of the electrode size because the electrode potential is independent of the electrode area in the biosensor. Importantly, because the substrate diffusion determines the enzyme reaction rate on the sensors, the dOCP/dt biosensors can essentially eliminate external influences such as temperature and pH. The result is the successful realization of a biosensor that is calibration‐free, making it a much more practical option.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Biosensing TechniquesPotentiometryCalibrationGlucoseElectrodesHydrogen-Ion ConcentrationLactic Acid

Stichwörter der Autor:innen

calibration‐freechemical sensorglucose sensorlactate sensorpotentiometric sensor

Substanzen

GlucoseLactic Acid

Autor:innen und Institutionen

  • Performance Chemicals Research, Kao Corporation, Wakayama, 640-8580, Japan.
  • Graduate School of Information Science and Technology, Hokkaido University, Sapporo, Hokkaido, 060-0814, Japan.

ORCID-Profile: Kuniharu Takei

Förderung

  • Japan Society for the Promotion of Science
  • AIP Network Laboratory

Zitieren

APA

Asai, M., Isshiki, N., Takesue, M., & Takei, K. (2025). Calibration‐Free and Highly Sensitive Potentiometric Enzyme‐Based Biosensors. Small Methods. https://doi.org/10.1002/smtd.202401010

Vancouver

Asai M, Isshiki N, Takesue M, Takei K. Calibration‐Free and Highly Sensitive Potentiometric Enzyme‐Based Biosensors. Small Methods. 2025. doi:10.1002/smtd.202401010

BibTeX
@article{Asai2025,
  author = {Asai, Mitsuo and Isshiki, Nobuyuki and Takesue, Masafumi and Takei, Kuniharu},
  title = {Calibration‐Free and Highly Sensitive Potentiometric Enzyme‐Based Biosensors},
  journal = {Small Methods},
  year = {2025},
  doi = {10.1002/smtd.202401010}
}