Safety and Efficacy of Direct Cryobiopsy Specimen Retrieval Through the 3.0‐mm Working Channel of a Therapeutic Bronchoscope Using a 1.1‐mm Cryoprobe: A Multicentre Prospective Observational Study

Yuki Takigawa iD, Masahide Oki, Ken Sato, Suzuka Matsuoka iD, Atsushi Torii iD, Jun Nishimura iD, Yuka Shinohara iD, Hiromi Watanabe, Masahiro Sano iD, Kenichiro Kudo, Akiko Sato, Yoshihito Kogure iD and 2 more

Respirology · 2026-09-16

DOI: 10.1002/resp.70314 ↗

At a glance

Journal article in Respirology, published online on 2026-09-16. Written by Yuki Takigawa and 13 co-authors. Institution: Department of Respiratory Medicine, NHO Okayama Medical Center, Okayama, Japan. Cited 1 times so far (Crossref); 15 references. The full text is freely available (open access, CC BY-NC 4.0).

AbstraktSubjectsBibliografické údajeCiteRelated studies

Abstract (original language)

Background and Objective Cryobiopsy using a 1.1‐mm cryoprobe through a therapeutic bronchoscope with a 3.0‐mm working channel permits en bloc retrieval of the cryoprobe and guide sheath while the bronchoscope remains wedged in the target bronchus, potentially facilitating immediate haemostatic control. This study investigated the safety and efficacy of this technique.

Methods

This prospective observational study enrolled patients with peripheral pulmonary lesions suspected of lung cancer who underwent forceps biopsy (FB), followed by cryobiopsy using a 1.1‐mm cryoprobe through a therapeutic bronchoscope with a 3.0‐mm working channel. The primary endpoint was procedural safety, as assessed by bleeding grade. The secondary endpoints included procedure‐related complications, diagnostic yield, specimen size, procedural success rate, and biomarker analysis success rate.

Results

Of 100 enrolled patients, 96 were included in the analysis, and 84 underwent cryobiopsy. Bleeding severity was Grade 0 in 35.4% (34 of 96), Grade 1 in 41.7% (40 of 96), Grade 2 in 8.3% (8 of 96), and Grade 3 in 2.1% (2 of 96) of patients. Diagnostic yields were 77.0% (74 of 96) for FB and 79.1% (76 of 96) for cryobiopsy, with an overall yield of 87.5% (84 of 96). Cryobiopsy provided the sole diagnosis in 10 patients and yielded significantly larger specimens than FB (7.06 vs. 2.46 mm 2 ; p < 0.001). The success rate of multiplex genetic testing was 95.8% for cryobiopsy (23 of 24).

Conclusion

Direct retrieval of cryobiopsy specimens through the working channel of a therapeutic bronchoscope is safe and feasible, yielding larger specimens with complementary diagnostic value.

Subjects

Author keywords

1.1‐mm cryoprobelung cancernon‐intubated cryobiopsypull through cryobiopsytherapeutic bronchoscopeultrathin cryobiopsy

Authors and institutions

  • Department of Respiratory Medicine, NHO Okayama Medical Center, Okayama, Japan.
  • Department of Respiratory Medicine, NHO Nagoya Medical Center, Nagoya, Japan.

ORCID profiles: Yuki Takigawa, Suzuka Matsuoka, Atsushi Torii, Jun Nishimura, Yuka Shinohara, Masahiro Sano, Yoshihito Kogure, Keiichi Fujiwara

Cite

APA

Takigawa, Y., Oki, M., Sato, K., Matsuoka, S., Torii, A., Nishimura, J., Shinohara, Y., Watanabe, H., Sano, M., Kudo, K., Sato, A., Kogure, Y., Fujiwara, K., & Shibayama, T. (2026). Safety and Efficacy of Direct Cryobiopsy Specimen Retrieval Through the 3.0‐mm Working Channel of a Therapeutic Bronchoscope Using a 1.1‐mm Cryoprobe: A Multicentre Prospective Observational Study. Respirology. https://doi.org/10.1002/resp.70314

Vancouver

Takigawa Y, Oki M, Sato K, Matsuoka S, Torii A, Nishimura J, et al. Safety and Efficacy of Direct Cryobiopsy Specimen Retrieval Through the 3.0‐mm Working Channel of a Therapeutic Bronchoscope Using a 1.1‐mm Cryoprobe: A Multicentre Prospective Observational Study. Respirology. 2026. doi:10.1002/resp.70314

BibTeX
@article{Takigawa2026,
  author = {Takigawa, Yuki and Oki, Masahide and Sato, Ken and Matsuoka, Suzuka and Torii, Atsushi and Nishimura, Jun and Shinohara, Yuka and Watanabe, Hiromi and Sano, Masahiro and Kudo, Kenichiro and Sato, Akiko and Kogure, Yoshihito and Fujiwara, Keiichi and Shibayama, Takuo},
  title = {Safety and Efficacy of Direct Cryobiopsy Specimen Retrieval Through the 3.0‐mm Working Channel of a Therapeutic Bronchoscope Using a 1.1‐mm Cryoprobe: A Multicentre Prospective Observational Study},
  journal = {Respirology},
  year = {2026},
  doi = {10.1002/resp.70314}
}
Start for free →