Low‐Complexity Robotic Device for Magnetic Resonance Imaging‐Guided Needle Biopsy
The International Journal of Medical Robotics and Computer Assisted Surgery · 2025
DOI: 10.1002/rcs.70121 ↗Kurz gefasst
Fachartikel in der Zeitschrift The International Journal of Medical Robotics and Computer Assisted Surgery, 2025. Verfasst von Anastasia Antoniou und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Electrical Engineering, Computer Engineering, and Informatics, Cyprus University of Technology, Limassol, Cyprus. Hauptthemen laut MEDLINE: Robotic Surgical Procedures, Image-Guided Biopsy, Magnetic Resonance Imaging, Robotics. Die Arbeit nennt 23 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
This work presents a low‐complexity robotic device for needle‐to‐target alignment designed to streamline Magnetic Resonance Imaging (MRI)‐guided biopsy procedures while maintaining physician control over insertion.
Methods
The robotic device was designed with a rigid frame incorporating two linear degrees of freedom for the in‐plane alignment of a needle guide with a predefined target. Trajectory planning and automated motion execution, including image transfer and sequence triggering, were managed through a custom MRI‐integrated software interface. Preliminary evaluation of alignment accuracy was performed in an agar phantom containing 5 and 10 mm tumour mimics.
Results
In phantom experiments, the system consistently achieved submillimeter alignment accuracy for both the 5‐mm and 10‐mm tumour models across all trials, without any operational failures.
Conclusions
These preliminary findings demonstrate the feasibility of the proposed robotic device for semiautomated MRI‐guided abdominal biopsy, pending further extensive preclinical testing.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Electrical Engineering, Computer Engineering, and Informatics, Cyprus University of Technology, Limassol, Cyprus.
- Diagnostic and Interventional Radiology Department, German Medical Institute, Limassol, Cyprus.
- Department of Electronics and Information Engineering, Hangzhou Dianzi University, Hangzhou, China.
ORCID-Profile: Anastasia Antoniou, Nikolas Evripidou, Christakis Damianou
Zitieren
APA
Antoniou, A., Evripidou, N., Georgiou, L., Chrysanthou, A., Christofi, A., Zhao, J., Yu, L., Li, W., & Damianou, C. (2025). Low‐Complexity Robotic Device for Magnetic Resonance Imaging‐Guided Needle Biopsy. The International Journal of Medical Robotics and Computer Assisted Surgery. https://doi.org/10.1002/rcs.70121
Vancouver
Antoniou A, Evripidou N, Georgiou L, Chrysanthou A, Christofi A, Zhao J, et al. Low‐Complexity Robotic Device for Magnetic Resonance Imaging‐Guided Needle Biopsy. The International Journal of Medical Robotics and Computer Assisted Surgery. 2025. doi:10.1002/rcs.70121
BibTeX
@article{Antoniou2025,
author = {Antoniou, Anastasia and Evripidou, Nikolas and Georgiou, Leonidas and Chrysanthou, Antreas and Christofi, Antonis and Zhao, Jufeng and Yu, Liyang and Li, Wenjun and Damianou, Christakis},
title = {Low‐Complexity Robotic Device for Magnetic Resonance Imaging‐Guided Needle Biopsy},
journal = {The International Journal of Medical Robotics and Computer Assisted Surgery},
year = {2025},
doi = {10.1002/rcs.70121}
}