Potential of Guidewire‐Clot Interaction Forces for Clot Detection in Robotic Mechanical Thrombectomy
The International Journal of Medical Robotics and Computer Assisted Surgery · 2025
DOI: 10.1002/rcs.70109 ↗Kurz gefasst
Fachartikel in der Zeitschrift The International Journal of Medical Robotics and Computer Assisted Surgery, 2025. Verfasst von Evgenia Roussinova und 6 weiteren Autor:innen. Beteiligte Einrichtung: REHAssist Group, Neuro-X Institute, EPFL, Lausanne, Switzerland. Hauptthemen laut MEDLINE: Thrombectomy, Robotic Surgical Procedures, Thrombosis. Die Arbeit nennt 34 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Swiss National Science Foundation.
Abstract (Originalsprache)
Background
Interventional neuroradiologists currently miss crucial information when facing an ischaemic stroke case as they cannot localise the thrombus precisely nor determine its mechanical properties, which are both important for the clot extraction procedure.
Methods
In this in vitro study, clot analogues of different stiffnesses and dimensions were inserted in an idealised arterial phantom using distinct insertion pressure levels. Guidewire‐clot interaction forces were recorded using a robotic device equipped with a force sensor. A model‐based clot detection method was developed.
Results
Clot stiffness, initial clot volume, and insertion pressure were all found to impact the guidewire‐clot interaction forces. In detecting the clot's proximal and distal edges, we achieved success rates of 86.5% and 71.2%, respectively, within a tolerance of 1.5 mm.
Conclusions
The study brings new perspectives for mechanical thrombectomy by demonstrating that multiple parameters influence the occlusion's mechanical state and by proposing algorithms for identifying the clot's extremities.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- REHAssist Group, Neuro-X Institute, EPFL, Lausanne, Switzerland.
- Division of Neuroradiology, Geneva University Hospitals, Geneva, Switzerland.
- Translational Neural Engineering Laboratory (TNE), Neuro-X Institute, EPFL, Lausanne, Switzerland.
- Biorobotics Institute and Department of Excellence in Robotics and AI, Scuola Superiore Sant'Anna, Pisa, Italy.
ORCID-Profile: Evgenia Roussinova
Förderung
- Swiss National Science Foundation
Zitieren
APA
Roussinova, E., Brina, O., Galand, W., Reymond, P., Micera, S., Machi, P., & Bouri, M. (2025). Potential of Guidewire‐Clot Interaction Forces for Clot Detection in Robotic Mechanical Thrombectomy. The International Journal of Medical Robotics and Computer Assisted Surgery. https://doi.org/10.1002/rcs.70109
Vancouver
Roussinova E, Brina O, Galand W, Reymond P, Micera S, Machi P, et al. Potential of Guidewire‐Clot Interaction Forces for Clot Detection in Robotic Mechanical Thrombectomy. The International Journal of Medical Robotics and Computer Assisted Surgery. 2025. doi:10.1002/rcs.70109
BibTeX
@article{Roussinova2025,
author = {Roussinova, Evgenia and Brina, Olivier and Galand, William and Reymond, Philippe and Micera, Silvestro and Machi, Paolo and Bouri, Mohamed},
title = {Potential of Guidewire‐Clot Interaction Forces for Clot Detection in Robotic Mechanical Thrombectomy},
journal = {The International Journal of Medical Robotics and Computer Assisted Surgery},
year = {2025},
doi = {10.1002/rcs.70109}
}
EUVIMED