Effect of Laminectomy Methods on the Surgical Safety of Automatic Laminectomy Robot

Ya Li iD, Da Lu, Xueqing Wu, Jiahao Lu, Yangyang Xu, Shijia Zhang, Le Zhang, Baoqing Pei

The International Journal of Medical Robotics and Computer Assisted Surgery · 2025

DOI: 10.1002/rcs.70031 ↗

Kurz gefasst

Fachartikel in der Zeitschrift The International Journal of Medical Robotics and Computer Assisted Surgery, 2025. Verfasst von Ya Li und 7 weiteren Autor:innen. Beteiligte Einrichtung: Beijing Key Laboratory for Design and Evaluation Technology of Advanced Implantable & Interventional Medical Devices, Beijing Advanced Innovation Cen…. Hauptthemen laut MEDLINE: Laminectomy, Robotic Surgical Procedures, Robotics. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 28 Literaturangaben. Gefördert von: Beijing Natural Science Foundation.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

The efficacy of laminectomy procedures is contingent on the method of resection. The objective of this study was to investigate the impact of different methods of resection on the surgical safety of automated laminectomy robots, an area that remains uncharted.

Methods

Lamina resection surgeries using both drilling and layer‐by‐layer methods, are performed on ovine spinal samples. An analysis of the force and lateral deviation at the end of the robotic arm is conducted.

Results

For the drilling and layer‐by‐layer lamina resection methods, the average peak force for the drilling method is 1.21 N, and deviations are within 1 mm. For the layer‐by‐layer resection method, the peak force reaches 2.39 N, and deviations are within 1.6 mm.

Conclusion

During the drilling resection, the ultrasonic osteotome experiences less force and minimal lateral deviation. This method demonstrates higher precision and safety in laminar resection surgeries. The drilling method should be the primary choice for robot manufacturers.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

LaminectomyRobotic Surgical ProceduresRoboticsAnimalsSheepEquipment DesignSpineAutomation

Stichwörter der Autor:innen

automatic laminectomydrilling and layer‐by‐layer methodforce feedback

Autor:innen und Institutionen

  • Beijing Key Laboratory for Design and Evaluation Technology of Advanced Implantable & Interventional Medical Devices, Beijing Advanced Innovation Center for Biomedical Engineering, School of Biological Science and Medic…
  • Tian Ji Laboratory, Beijing Tinavi Medical Technology Co., Ltd, Beijing, China.

ORCID-Profile: Ya Li

Förderung

  • Beijing Natural Science Foundation

Zitieren

APA

Li, Y., Lu, D., Wu, X., Lu, J., Xu, Y., Zhang, S., Zhang, L., & Pei, B. (2025). Effect of Laminectomy Methods on the Surgical Safety of Automatic Laminectomy Robot. The International Journal of Medical Robotics and Computer Assisted Surgery. https://doi.org/10.1002/rcs.70031

Vancouver

Li Y, Lu D, Wu X, Lu J, Xu Y, Zhang S, et al. Effect of Laminectomy Methods on the Surgical Safety of Automatic Laminectomy Robot. The International Journal of Medical Robotics and Computer Assisted Surgery. 2025. doi:10.1002/rcs.70031

BibTeX
@article{Li2025,
  author = {Li, Ya and Lu, Da and Wu, Xueqing and Lu, Jiahao and Xu, Yangyang and Zhang, Shijia and Zhang, Le and Pei, Baoqing},
  title = {Effect of Laminectomy Methods on the Surgical Safety of Automatic Laminectomy Robot},
  journal = {The International Journal of Medical Robotics and Computer Assisted Surgery},
  year = {2025},
  doi = {10.1002/rcs.70031}
}
Kostenlos starten →