Development of the headspace gas chromatography–tandem mass spectrometry method for ethylene oxide and ethylene chlorohydrin residue in medical devices

Yi Chun Chiang iD, Ming Chih Fang iD, Chuan Kai Yang, Sheng Wei Wang, Su Hsiang Tseng

Rapid Communications in Mass Spectrometry · 2024

DOI: 10.1002/rcm.9869 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Rapid Communications in Mass Spectrometry, 2024. Verfasst von Yi Chun Chiang und 4 weiteren Autor:innen. Beteiligte Einrichtung: Division of Research and Analysis, Taiwan Food and Drug Administration, Ministry of Health and Welfare, Taipei City, Taiwan. Hauptthemen laut MEDLINE: Ethylene Oxide, Tandem Mass Spectrometry, Gas Chromatography-Mass Spectrometry. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 18 Literaturangaben. Gefördert von: Food and Drug Administration, Ministry of Health and Welfare of Taiwan.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Rationale

Ethylene oxide (EO) sterilization is commonly employed for the sterilization of medical devices and has a very high market share. However, EO and its metabolite ethylene chlorohydrin (ECH) are toxic to humans. In compliance with the classification and residue limits of medical devices defined by ISO 10993‐7, our study established two extraction methods for the testing of EO and ECH.

Methods

The first method involves simulated‐use extraction using water as the extraction solvent. While the second, exhaustive extraction, directly extracts sample through headspace sampling analysis. Gas chromatography–tandem mass spectrometry in multiple reaction monitoring mode was utilized, requiring only 16 min. Then, the developed method was applied to assess 10 commercially available medical devices sterilized by EO.

Results

In simulated‐use extraction, calibration curves were evaluated in the range of 1–100 and 5–500 μg for EO and ECH, respectively ( r > 0.999). Inter‐day recoveries ranged from 85.0% to 95.2% and from 94.8% to 102.4%. In exhaustive extraction, calibration curves spanned 0.5–50 and 2–200 μg for EO and ECH, respectively ( r > 0.999). Inter‐day recoveries ranged from 101.6% to 102.1% for EO and from 98.1% to 102.2% for ECH. After analysis of the 10 commercially available medical devices, two cotton swabs were found to have ECH of 35.1 and 28.4 μg per device, and four medical devices were found to have EO with concentration below the limit of quantification. Meanwhile, we found that the EO internal standard (propylene oxide) recommended by ISO 10993‐7 had interference problems with other similar substances and was not suitable as an internal standard for EO.

Conclusions

This study offers a sensitive and straightforward analytical approach to EO and ECH residues in a variety of medical devices. In addition, the results show that the EO or ECH content of these types of medical devices in our study falls below the regulatory limits, therefore instilling confidence among consumers regarding their safe use.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Ethylene OxideTandem Mass SpectrometryGas Chromatography-Mass SpectrometryEquipment and SuppliesLimit of DetectionEthylenesReproducibility of ResultsEquipment ContaminationSterilization

Substanzen

Ethylene OxideEthylenes

Autor:innen und Institutionen

  • Division of Research and Analysis, Taiwan Food and Drug Administration, Ministry of Health and Welfare, Taipei City, Taiwan.
  • Department of Food Science, National Taiwan Ocean University, Keelung, Taiwan.

ORCID-Profile: Yi Chun Chiang, Ming Chih Fang

Förderung

  • Food and Drug Administration, Ministry of Health and Welfare of Taiwan

Zitieren

APA

Chiang, Y. C., Fang, M. C., Yang, C. K., Wang, S. W., & Tseng, S. H. (2024). Development of the headspace gas chromatography–tandem mass spectrometry method for ethylene oxide and ethylene chlorohydrin residue in medical devices. Rapid Communications in Mass Spectrometry. https://doi.org/10.1002/rcm.9869

Vancouver

Chiang YC, Fang MC, Yang CK, Wang SW, Tseng SH. Development of the headspace gas chromatography–tandem mass spectrometry method for ethylene oxide and ethylene chlorohydrin residue in medical devices. Rapid Communications in Mass Spectrometry. 2024. doi:10.1002/rcm.9869

BibTeX
@article{Chiang2024,
  author = {Chiang, Yi Chun and Fang, Ming Chih and Yang, Chuan Kai and Wang, Sheng Wei and Tseng, Su Hsiang},
  title = {Development of the headspace gas chromatography–tandem mass spectrometry method for ethylene oxide and ethylene chlorohydrin residue in medical devices},
  journal = {Rapid Communications in Mass Spectrometry},
  year = {2024},
  doi = {10.1002/rcm.9869}
}
Kostenlos starten →