EUVIMEDEUVIMED entdecken
Crossref

Development of an arrow solid-phase microextraction GC–MS/MS method for the determination of 69 odorants in water

FeiFei Dai, Ruoyu Zhang, Dandan Nie, Shuqing Ma, Yurong Qin

Journal of Chromatography A · 2026

DOI: 10.1016/j.chroma.2026.466703 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Chromatography A, 2026. Verfasst von FeiFei Dai und 4 weiteren Autor:innen. Beteiligte Einrichtung: Weifang Municipal Center for Disease Control and Prevention, Weifang, Shandong, 261061, China. Hauptthemen laut MEDLINE: Gas Chromatography-Mass Spectrometry, Solid Phase Microextraction, Odorants, Water Pollutants, Chemical. Die Arbeit nennt 17 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

A quantitative method for determining 69 odorants in water was developed using arrow-type headspace solid-phase microextraction (SPME) coupled with gas chromatography-tandem mass spectrometry (GC-MS/MS). Key extraction parameters, including extraction and desorption temperatures and times, NaCl addition, fiber type, and rotation speed, were systematically optimized. Under the optimized conditions, target compounds were efficiently enriched, separated on a polar chromatographic column, detected in selected reaction monitoring (SRM) mode, and quantified using external calibration. The method demonstrated high sensitivity, with limits of detection (LODs) ranging from 0.015 to 75 ng/L, and 55 compounds exhibiting LODs below 10 ng/L. All analytes showed excellent linearity (R2 > 0.995) within their respective ranges. Precision was satisfactory, with relative standard deviations (RSDs) between 1.10 % and 17.85 %, including 59 compounds with RSDs

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Gas Chromatography-Mass SpectrometrySolid Phase MicroextractionOdorantsWater Pollutants, ChemicalTandem Mass SpectrometryLimit of DetectionReproducibility of ResultsWaterLinear Models

Stichwörter der Autor:innen

Gas chromatography-tandem mass spectrometryOdor substancesSPME arrowWater

Substanzen

Water Pollutants, ChemicalWater

Autor:innen und Institutionen

  • Weifang Municipal Center for Disease Control and Prevention, Weifang, Shandong, 261061, China.
  • School of Public Health, Shandong Second Medical University, Weifang, Shandong, 261053, China. Electronic address: ruoyu168@126.com.
  • Weifang Municipal Center for Disease Control and Prevention, Weifang, Shandong, 261061, China. Electronic address: niedan777@sohu.com.
  • Thermo Fisher Scientific, Beijing World Trade Center, Beijing, 100102, China.

Zitieren

APA

Dai, F., Zhang, R., Nie, D., Ma, S., & Qin, Y. (2026). Development of an arrow solid-phase microextraction GC–MS/MS method for the determination of 69 odorants in water. Journal of Chromatography A. https://doi.org/10.1016/j.chroma.2026.466703

Vancouver

Dai F, Zhang R, Nie D, Ma S, Qin Y. Development of an arrow solid-phase microextraction GC–MS/MS method for the determination of 69 odorants in water. Journal of Chromatography A. 2026. doi:10.1016/j.chroma.2026.466703

BibTeX
@article{Dai2026,
  author = {Dai, FeiFei and Zhang, Ruoyu and Nie, Dandan and Ma, Shuqing and Qin, Yurong},
  title = {Development of an arrow solid-phase microextraction GC–MS/MS method for the determination of 69 odorants in water},
  journal = {Journal of Chromatography A},
  year = {2026},
  doi = {10.1016/j.chroma.2026.466703}
}
Kostenlos starten →