Flow‐Driver‐Generated Synchronized Nasal Intermittent Positive‐Pressure Ventilation Versus Biphasic Positive Airway Pressure After Extubation in Preterm Infants
Pediatric Pulmonology · 2025
DOI: 10.1002/ppul.71410 ↗At a glance
Beobachtungsstudie in Pediatric Pulmonology, 2025. Written by Toshihiko Suzuki and 7 co-authors. Institution: Center for Maternal-Neonatal Care, Division of Neonatology, Nagoya University Hospital, Nagoya, Aichi, Japan. Major MEDLINE topics: Airway Extubation, Intermittent Positive-Pressure Ventilation, Continuous Positive Airway Pressure, Respiratory Distress Syndrome, Newborn. 40 references. The full text is freely available (open access, CC BY-NC 4.0). Funded by: This work was supported by a grant from the Morinaga Foundation for Health & Nutrition..
Topics: Neonatologie, Beatmung und Weaning
Abstract (original language)
Objective
To compare the efficacy and safety of flow‐driver‐generated synchronized nasal intermittent positive‐pressure ventilation (SNIPPV) and biphasic positive airway pressure (BiPAP) after extubation in preterm infants. Hypothesis SNIPPV delivered via a flow‐driver device improves post‐extubation outcomes compared with non‐synchronized BiPAP. Study Design This single‐center, retrospective, observational cohort study was conducted in the neonatal intensive care unit of Nagoya University Hospital, Nagoya, Japan. Patient Selection Preterm infants born at < 34 weeks of gestation who underwent endotracheal ventilation after birth and were subsequently supported with either BiPAP (from October 2017 to March 2020) or SNIPPV (from April 2020 to September 2022) after extubation. Infants with major anomalies or surgical ventilator dependence were excluded. Methodology Data on demographic and clinical characteristics, respiratory outcomes, and complications were collected. The primary outcome was reintubation within 72 h. Secondary outcomes were bronchopulmonary dysplasia (BPD), respiratory support duration, and major complications.
Results
Sixty‐seven infants were included (BiPAP n = 37, SNIPPV n = 30). Reintubation within 72 h occurred in 6.7% and 29.7% of infants in the SNIPPV and BiPAP groups, respectively ( p = 0.028). The incidence of respiratory distress‐related failure was significantly lower in the SNIPPV group (3.3% vs. 24.3%, p = 0.019). There were no significant differences in BPD at 36 or 40 weeks, oxygen therapy duration, or complications such as necrotizing enterocolitis, retinopathy of prematurity, or severe neurological injury.
Conclusion
Flow‐driver‐generated SNIPPV significantly reduces early extubation failure in preterm infants compared to BiPAP without increasing the rate of adverse outcomes. This is a practical and effective alternative to noninvasive respiratory support.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- Center for Maternal-Neonatal Care, Division of Neonatology, Nagoya University Hospital, Nagoya, Aichi, Japan.
ORCID profiles: Toshihiko Suzuki, Kazuto Ueda, Akinobu Taniguchi, Takashi Maeda, Ryuichi Tanaka, Ryosuke Miura, Yukako Muramatsu, Yoshiaki Sato
Funding
- This work was supported by a grant from the Morinaga Foundation for Health & Nutrition.
Cite
APA
Suzuki, T., Ueda, K., Taniguchi, A., Maeda, T., Tanaka, R., Miura, R., Muramatsu, Y., & Sato, Y. (2025). Flow‐Driver‐Generated Synchronized Nasal Intermittent Positive‐Pressure Ventilation Versus Biphasic Positive Airway Pressure After Extubation in Preterm Infants. Pediatric Pulmonology. https://doi.org/10.1002/ppul.71410
Vancouver
Suzuki T, Ueda K, Taniguchi A, Maeda T, Tanaka R, Miura R, et al. Flow‐Driver‐Generated Synchronized Nasal Intermittent Positive‐Pressure Ventilation Versus Biphasic Positive Airway Pressure After Extubation in Preterm Infants. Pediatric Pulmonology. 2025. doi:10.1002/ppul.71410
BibTeX
@article{Suzuki2025,
author = {Suzuki, Toshihiko and Ueda, Kazuto and Taniguchi, Akinobu and Maeda, Takashi and Tanaka, Ryuichi and Miura, Ryosuke and Muramatsu, Yukako and Sato, Yoshiaki},
title = {Flow‐Driver‐Generated Synchronized Nasal Intermittent Positive‐Pressure Ventilation Versus Biphasic Positive Airway Pressure After Extubation in Preterm Infants},
journal = {Pediatric Pulmonology},
year = {2025},
doi = {10.1002/ppul.71410}
}
EUVIMED