Fachartikel in der Zeitschrift Pediatric Pulmonology, 2025. Verfasst von Koen P Dijkman und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Neonatology, Máxima MC, Veldhoven, The Netherlands. Hauptthemen laut MEDLINE: Infant, Extremely Premature, Oxygen Inhalation Therapy, Oxygen Saturation, Oxygen. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: The authors received no specific funding for this work..

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

Automated or closed‐loop FiO2‐control (FiO2‐c) has been developed to maintain oxygen saturation (SpO2) within a target range more effectively. However, knowledge of the effects of prolonged use of FiO2‐c in extremely preterm infants during a more integral part of NICU admission is limited.

Methods

Twenty five extremely preterm infants (gestational age [GA] < 28 weeks) who survived until NICU discharge andreceived FiO2‐c by Predictive Intelligent Control of Oxygenation (PRICO) were matched 1:1 for GA, birthweight, sex andsurvival to a cohort receiving routine manual FiO2 control. The proportions of time within the SpO2 target range, (severe) hypoxia, hyperoxia, FiO2 and SpO2 for all days and days on supplemental oxygen, for both the whole period and per week, were compared.

Results

Infants in the FiO2‐c cohort received the intervention during 98% of the 7‐week study period. Overall, with FiO2‐c, a small, non‐significant increase in time within SpO2 target range was observed: mean difference 0.5% (95% CI [−5.0, 6.0]). However, when requiring supplemental oxygen during the first 2 weeks of life, time within SpO2 target range significantly increased with FiO2‐c, while time in hyperoxia decreased: mean differences in Week 1: 9.9% (95% CI [3.1, 16.7]) and −10.2% (95% CI [−17.1, −3.3]); in Week 2: 9.5% (95% CI [1.4, 17.6]) and −9.9% (95% CI [−19.2, −0.2]).

Conclusion

Despite limited overall effect, continuous use of FiO2‐c in extremely preterm infants requiring supplemental oxygen was associated with an increased time within the SpO2 target range during the first 2 weeks of life, a critical window for hyperoxia‐related diseases.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Infant, Extremely PrematureOxygen Inhalation TherapyOxygen SaturationOxygenHumansInfant, NewbornMaleFemaleHyperoxiaCohort StudiesIntensive Care Units, NeonatalHypoxiaGestational AgeOximetry

Stichwörter der Autor:innen

automated oxygen controlclosed loop oxygen controlhyperoxiahypoxiapreterm

Substanzen

Oxygen

Autor:innen und Institutionen

  • Department of Neonatology, Máxima MC, Veldhoven, The Netherlands.
  • Department of Applied Physics, TU/e Eindhoven, University of Technology, Eindhoven, The Netherlands.
  • MMC Science Office, Máxima MC, Veldhoven, The Netherlands.
  • Department of Clinical Physics, Máxima MC, Veldhoven, The Netherlands.
  • Department of Neonatal and Pediatric Intensive Care, Division of Neonatology, Erasmus MC, Rotterdam, The Netherlands.
  • Department of Neonatology and Centre for Paediatric Clinical Studies, University Children's Hospital Tübingen, Tübingen, Baden-Württemberg, Germany.

ORCID-Profile: Koen P Dijkman, Hendrik J Niemarkt

Förderung

  • The authors received no specific funding for this work.

Zitieren

APA

Dijkman, K. P., Delbressine, J. J., Dieleman, J. P., Mohns, T., Andriessen, P., Pul, C. V., Reiss, I. K. M., Franz, A. R., & Niemarkt, H. J. (2025). Continuous Application of Closed‐Loop FiO2‐Control in Extremely Preterm Infants: A Matched Cohort Single‐Center Study. Pediatric Pulmonology. https://doi.org/10.1002/ppul.71122

Vancouver

Dijkman KP, Delbressine JJ, Dieleman JP, Mohns T, Andriessen P, Pul CV, et al. Continuous Application of Closed‐Loop FiO2‐Control in Extremely Preterm Infants: A Matched Cohort Single‐Center Study. Pediatric Pulmonology. 2025. doi:10.1002/ppul.71122

BibTeX
@article{Dijkman2025,
  author = {Dijkman, Koen P and Delbressine, Jesse J and Dieleman, Jeanne P and Mohns, Thilo and Andriessen, Peter and Pul, Carola van and Reiss, Irwin K M and Franz, Axel R and Niemarkt, Hendrik J},
  title = {Continuous Application of Closed‐Loop FiO2‐Control in Extremely Preterm Infants: A Matched Cohort Single‐Center Study},
  journal = {Pediatric Pulmonology},
  year = {2025},
  doi = {10.1002/ppul.71122}
}
Kostenlos starten →