Performance and safety of volume guarantee ventilation in neonates using the fabian ventilator, a multicenter study
Pediatric Pulmonology · 2024
DOI: 10.1002/ppul.27264 ↗Kurz gefasst
Multizentrische Studie in der Zeitschrift Pediatric Pulmonology, 2024. Verfasst von Zoltan Molnar und 10 weiteren Autor:innen. Beteiligte Einrichtung: Neonatal Intensive Care Unit, Oxford University Hospitals NHS Trust, Oxford, UK. Hauptthemen laut MEDLINE: Tidal Volume, Ventilators, Mechanical, Respiration, Artificial, Intensive Care Units, Neonatal. Die Arbeit nennt 16 Literaturangaben. Gefördert von: Vyaire Medical.
Abstract (Originalsprache)
Objective
To evaluate the performance (i.e., agreement between set and measured parameters) and safety (adverse events, device malfunctions, and ventilator alarms) of the fabian HFOi neonatal ventilator in volume guaranteed (VG) mode during conventional ventilation. To analyze the impact of leakage around the endotracheal tube and the set maximum allowed inflating pressure (Pmax).
Design
Prospective multicenter observational study.
Methods
Clinical and ventilator data were collected from 71 infants receiving VG ventilation for ≥12 h in four neonatal intensive care units (NICUs). Ventilator settings, parameters, and alarms were downloaded with 0.5 Hz sampling rate.
Results
Data from 4,341 h of ventilation were analyzed. The median (interquartile range, IQR) of the absolute difference between the target and measured expired tidal volume was 0.76 (0.51–1.16) mL/kg. It was less when leak was
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Neonatal Intensive Care Unit, Oxford University Hospitals NHS Trust, Oxford, UK.
- Department of Neonatology, Emma Children's Hospital, Amsterdam UMC, Amsterdam, The Netherlands.
- Amsterdam Reproduction & Development Research Institute, Amsterdam, The Netherlands.
- University hospital Centre of Postgraduate Medical Education, SPSK, W. Orłowskiego, Warsaw, Poland.
- II Department of Neonatology, Neonatal Biophysical Monitoring and Cardiopulmonary Therapies Research Unit, Poznań University of Medical Sciences Poznan, Poland.
- Neonatal Intensive Care Unit, Cambridge University Hospitals NHS Trust, Cambridge, UK.
ORCID-Profile: Ruud W van Leuteren, Katarzyna Wróblewska-Seniuk, Gusztav Belteki
Förderung
- Vyaire Medical
Zitieren
APA
Molnar, Z., van Leuteren, R. W., Wilińka, M., Szczapa, T., Hutten, J., Gupta, A., Piwowarczyk, P., Wróblewska-Seniuk, K., Chojnacka, K., van Kaam, A. H., & Belteki, G. (2024). Performance and safety of volume guarantee ventilation in neonates using the fabian ventilator, a multicenter study. Pediatric Pulmonology. https://doi.org/10.1002/ppul.27264
Vancouver
Molnar Z, van Leuteren RW, Wilińka M, Szczapa T, Hutten J, Gupta A, et al. Performance and safety of volume guarantee ventilation in neonates using the fabian ventilator, a multicenter study. Pediatric Pulmonology. 2024. doi:10.1002/ppul.27264
BibTeX
@article{Molnar2024,
author = {Molnar, Zoltan and van Leuteren, Ruud W and Wilińka, Maria and Szczapa, Tomasz and Hutten, Jeroen and Gupta, Amit and Piwowarczyk, Pawel and Wróblewska-Seniuk, Katarzyna and Chojnacka, Karolina and van Kaam, Anton H and Belteki, Gusztav},
title = {Performance and safety of volume guarantee ventilation in neonates using the fabian ventilator, a multicenter study},
journal = {Pediatric Pulmonology},
year = {2024},
doi = {10.1002/ppul.27264}
}
EUVIMED