Real‐Time Measurement of Particulate Concentration: Laser Surgery With Jet Ventilation

S Syam iD, Justin Jou, Bryan E Schmidt iD, N Scott Howard iD

OTO Open · 2026-10 · 10 (4)

DOI: 10.1002/oto2.70299 ↗

Kurz gefasst

Fachartikel in der Zeitschrift OTO Open, online erschienen am 15.11.2026. Verfasst von S Syam und 3 weiteren Autor:innen. Beteiligte Einrichtung: Mechanical and Aerospace Engineering Case Western Reserve University Cleveland Ohio USA. Die Arbeit nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Institute of Biomedical Imaging and Bioengineering.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

To quantify the particulate concentration in exhalation plumes during laser surgery procedures under jet ventilation and to examine the effect of mitigation strategies via suction. Study Design In vivo quantitative study.

Setting

Tertiary care operating room.

Methods

The particulate concentration in the jet ventilation plume was measured in real‐time using an optical particle sizer. Several suction configurations were tested to evaluate their effects on the particle concentrations in the plume.

Results

The total particle concentration increased significantly during CO 2 laser use with simultaneous jet ventilation. The sustained post‐procedure total concentration was on average 27% higher than the pre‐procedure level. With the application of two sources of suction, total concentrations of small, medium, and larger particles in the exhalation plume during CO 2 laser ablation were reduced by 54%, 85%, and 89%, respectively, compared to no suction. During the recovery phase, the reductions were 97%, 94%, and 84%, respectively. Although suction reduced the concentration of all particle sizes in the plume, the concentration of submicron particles remained high.

Conclusion

Suction significantly reduced particle concentration in the plume during laser ablation compared to the appreciable levels produced with no suction; the two‐suction use case was found to be most effective. However, the elimination of submicron particulates in the plume by suction devices was found to be less efficient. We recommend using one or two suction devices and high‐filtration masks during jet ventilation‐assisted laser procedures to minimize the risk posed by the aerosols generated during these procedures. Further mitigation studies are warranted.

Schlagwörter

Stichwörter der Autor:innen

aerosol generating proceduresjet ventilationlaser surgeryparticulate concentrationsurgical plume

Autor:innen und Institutionen

  • Mechanical and Aerospace Engineering Case Western Reserve University Cleveland Ohio USA.
  • School of Medicine, Otolaryngology, Head and Neck Surgery University Hospitals, Cleveland Medical Center Ohio USA.

ORCID-Profile: S Syam, Bryan E Schmidt, N Scott Howard

Förderung

  • National Institute of Biomedical Imaging and Bioengineering

Zitieren

APA

Syam, S., Jou, J., Schmidt, B. E., & Howard, N. S. (2026). Real‐Time Measurement of Particulate Concentration: Laser Surgery With Jet Ventilation. OTO Open, 10(4). https://doi.org/10.1002/oto2.70299

Vancouver

Syam S, Jou J, Schmidt BE, Howard NS. Real‐Time Measurement of Particulate Concentration: Laser Surgery With Jet Ventilation. OTO Open. 2026;10(4). doi:10.1002/oto2.70299

BibTeX
@article{Syam2026,
  author = {Syam, S and Jou, Justin and Schmidt, Bryan E and Howard, N Scott},
  title = {Real‐Time Measurement of Particulate Concentration: Laser Surgery With Jet Ventilation},
  journal = {OTO Open},
  year = {2026},
  volume = {10},
  number = {4},
  doi = {10.1002/oto2.70299}
}
Kostenlos starten →