Journal article in Otolaryngology–Head and Neck Surgery, 2024. Written by Jeong-Seo Kim and 2 co-authors. Institution: Hearing Research Laboratory, Samsung Medical Center, Seoul, South Korea. Major MEDLINE topics: Cochlear Implants, Prosthesis Design, Action Potentials. Cited 3 times so far (Crossref); 15 references.

AstrattSubjectsDettalji biblijografiċiCiteRelated studies

Abstract (original language)

Objective

Cochlear implant (CI) electrode design has changed over time. Changes in intracochlear electrode design might influence the spread of neural activation along the auditory nerve and the number of independent channels. This study aimed to investigate the impact of intracochlear electrode design on the electrode‐neuron interface using electrophysiological measures. Study Design Prospective cohort study.

Setting

A single tertiary hospital.

Methods

Fifty‐two ears who were implanted with CI divided into 3 groups based on the design of intracochlear electrode arrays. Twenty‐three ears were implanted with lateral wall straight electrodes. Eighteen ears were implanted with the slim perimodiolar electrode, and 11 ears were implanted with the old perimodiolar electrode. Various electrically‐evoked compound action potential (ECAP) metrics were measured to quantify spread of excitation and channel interaction.

Results

ECAP threshold and slope were not significantly different among groups. ECAP spread of excitation (SOE) half‐width and channel interaction index (CII) were significantly larger in subjects implanted with the lateral wall straight electrodes, indicating a wider spread of excitation compared to those with perimodiolar electrodes. Electrode impedance was significantly lower in subjects implanted with perimodiolar electrodes than those with lateral wall electrodes.

Conclusion

Perimodiolar electrode groups yielded significantly narrower SOE half‐widths and smaller CII than the lateral wall straight electrode group. This may indicate that the electrode array that hugged the modiolus had less overlap in neural excitation between adjacent electrodes, resulting in reduced channel interaction and potentially better spectral resolution than the electrode array positioned more laterally.

Subjects

MeSH terms (MEDLINE)

Cochlear ImplantsProsthesis DesignAction PotentialsHumansProspective StudiesMaleFemaleMiddle AgedAdultAgedCochlear ImplantationEvoked Potentials, AuditoryElectrodes, Implanted

Author keywords

channel interactioncochlear implantelectrically‐evoked compound action potential (ECAP)intracochlear electrode arrayspread of excitation

Authors and institutions

  • Hearing Research Laboratory, Samsung Medical Center, Seoul, South Korea.
  • Medical Research Institute, Sungkyunkwan University, Suwon, South Korea.
  • Department of Speech-Language Pathology and Aural Rehabilitation, Woosong University, Daejeon, South Korea.
  • Soree Ear Clinic, Seoul, South Korea.
  • Department of Otolaryngology-Head and Neck Surgery, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, South Korea.

ORCID profiles: Jeong-Seo Kim, Sung Hwa Hong, Il Joon Moon

Cite

APA

Kim, J. S., Hong, S. H., & Moon, I. J. (2024). Association Between Intracochlear Electrode Design and Electrically‐Evoked Compound Action Potential Measures in Cochlear Implant Users. Otolaryngology–Head and Neck Surgery. https://doi.org/10.1002/ohn.799

Vancouver

Kim JS, Hong SH, Moon IJ. Association Between Intracochlear Electrode Design and Electrically‐Evoked Compound Action Potential Measures in Cochlear Implant Users. Otolaryngology–Head and Neck Surgery. 2024. doi:10.1002/ohn.799

BibTeX
@article{Kim2024,
  author = {Kim, Jeong-Seo and Hong, Sung Hwa and Moon, Il Joon},
  title = {Association Between Intracochlear Electrode Design and Electrically‐Evoked Compound Action Potential Measures in Cochlear Implant Users},
  journal = {Otolaryngology–Head and Neck Surgery},
  year = {2024},
  doi = {10.1002/ohn.799}
}
Start for free →