Complications of Novel Radiofrequency Device Use in Rhinology: A MAUDE Analysis
Otolaryngology–Head and Neck Surgery · 2024-02 · 170 (2) · 605-609
DOI: 10.1002/ohn.536 ↗At a glance
Journal article in Otolaryngology–Head and Neck Surgery, published online on 2023-09-25. Written by Sina J Torabi and 5 co-authors. Institution: Department of Otolaryngology-Head and Neck Surgery, University of California, Irvine, Orange, California, USA. Major MEDLINE topics: Necrosis. Cited 9 times so far (Crossref); 8 references.
Abstract (original language)
With the widespread adoption of intranasal radiofrequency (RF) devices, our objective was to report national adverse events (AEs) associated with their use. The Food and Drug Administration's Manufacturer and User Facility Device Experience was queried. A total of 24 device‐related AEs were reported, 11 (45.8%) for Celon® (Olympus), 3 (12.5%) for Vivaer® (Aerin), 2 (8.3%) for Neuromark® (Neurent), and 8 (33.3%) for Rhinaer® (Aerin). Seven (63.6%) of the Celon®‐related complications were related to tissue necrosis (largely user error‐related), but 1 (9.1%) episode of pediatric ocular palsy was also reported. Vivaer® complications included synechiae formation, a mucosal perforation, and a case of empty nose syndrome. Of the posterior nasal nerve ablating devices, 9 of 10 AEs were epistaxes, of which 7 (77.8%) required operative intervention. Surgeons should exercise vigilance and tissue‐appropriate device settings when utilizing RF devices. Epistaxis and tissue necrosis may occur, as well as more rare, but devastating, complications.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- Department of Otolaryngology-Head and Neck Surgery, University of California, Irvine, Orange, California, USA.
ORCID profiles: Sina J Torabi, Benjamin F Bitner, Eric H Abello
Cite
APA
Torabi, S. J., Bitner, B. F., Abello, E. H., Nguyen, T. V., Wong, B. J. F., & Kuan, E. C. (2024). Complications of Novel Radiofrequency Device Use in Rhinology: A MAUDE Analysis. Otolaryngology–Head and Neck Surgery, 170(2), 605-609. https://doi.org/10.1002/ohn.536
Vancouver
Torabi SJ, Bitner BF, Abello EH, Nguyen TV, Wong BJF, Kuan EC. Complications of Novel Radiofrequency Device Use in Rhinology: A MAUDE Analysis. Otolaryngology–Head and Neck Surgery. 2024;170(2):605-609. doi:10.1002/ohn.536
BibTeX
@article{Torabi2024,
author = {Torabi, Sina J and Bitner, Benjamin F and Abello, Eric H and Nguyen, Theodore V and Wong, Brian J F and Kuan, Edward C},
title = {Complications of Novel Radiofrequency Device Use in Rhinology: A MAUDE Analysis},
journal = {Otolaryngology–Head and Neck Surgery},
year = {2024},
volume = {170},
number = {2},
pages = {605--609},
doi = {10.1002/ohn.536}
}
EUVIMED