EUVIMEDEUVIMED entdecken
Crossref

Reliability of MScanFit Motor Unit Number Estimation in the Trapezius Muscle

Daniel Mosgaard Sørensen, Hatice Tankisi iD

Muscle & Nerve · 2025

DOI: 10.1002/mus.28303 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Muscle & Nerve, 2025. Verfasst von Daniel Mosgaard Sørensen und 1 weiteren Autor:innen. Beteiligte Einrichtung: Department of Clinical Neurophysiology, Aarhus University Hospital, Aarhus, Denmark. Hauptthemen laut MEDLINE: Superficial Back Muscles, Electromyography, Action Potentials. Die Arbeit nennt 26 Literaturangaben. Gefördert von: Aage & Johanne Louis-Hansens Foundation, Grosserer L. F. Foghts Foundation, Dagmar Marshalls Fond.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction/Aims MScanFit motor unit number estimation (MUNE) is the most recent MUNE method which has shown promising results in extremity muscles, but it has not been applied to bulbar muscles. Bulbar muscles are particularly important in the diagnosis of amyotrophic lateral sclerosis (ALS). This study aimed to investigate the feasibility and reliability of MScanFit MUNE in the accessory nerve and trapezius muscles.

Methods

A total of twenty healthy participants were examined twice within 1–2 weeks. We extracted the MScanFit MUNE and size parameter, and compound muscle action potential (CMAP) amplitude values. The reliability of these parameters was assessed using the intra‐rater coefficient of variation (CoV), intraclass correlation coefficient (ICC), and Bland–Altman plots. We also correlated MUNE values with CMAP amplitudes using correlation coefficients.

Results

Mean MUNE values (Day 1 = 132.1 and Day 2 = 137.4), CMAP amplitudes (Day 1 = 9.71 mV and Day 2 = 10.10 mV) and size parameters did not differ between the two sessions ( p > 0.05). CoV showed excellent reliability for MUNE values, size parameters, and CMAP amplitudes (CoV < 7%) whereas ICCs showed moderate reliability for MUNE values (ICC = 0.619), poor to moderate reliability for size parameters (between 0.393 and 0.689), and good reliability for CMAP amplitude (ICC = 0.864) There was no correlation between MUNE values and CMAP amplitudes.

Discussion

MScanFit MUNE is applicable and mostly reliable in the trapezius muscle. Further studies in patients are needed to investigate the sensitivity of MScanFit in this muscle in detecting motor unit loss, particularly in ALS.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Superficial Back MusclesElectromyographyAction PotentialsHumansMaleFemaleReproducibility of ResultsAdultMotor NeuronsYoung AdultMiddle AgedRecruitment, NeurophysiologicalAmyotrophic Lateral Sclerosis

Stichwörter der Autor:innen

MScanFit MUNEaccessory nervebulbar innervated musclemotor unit number estimationtrapezius superior muscle

Autor:innen und Institutionen

  • Department of Clinical Neurophysiology, Aarhus University Hospital, Aarhus, Denmark.
  • Department of Clinical Medicine, Aarhus University, Aarhus, Denmark.

ORCID-Profile: Hatice Tankisi

Förderung

  • Aage & Johanne Louis-Hansens Foundation
  • Grosserer L. F. Foghts Foundation
  • Dagmar Marshalls Fond
  • The Jascha Foundation,
  • Lundbeck Foundation

Zitieren

APA

Sørensen, D. M., & Tankisi, H. (2025). Reliability of MScanFit Motor Unit Number Estimation in the Trapezius Muscle. Muscle & Nerve. https://doi.org/10.1002/mus.28303

Vancouver

Sørensen DM, Tankisi H. Reliability of MScanFit Motor Unit Number Estimation in the Trapezius Muscle. Muscle & Nerve. 2025. doi:10.1002/mus.28303

BibTeX
@article{Srensen2025,
  author = {Sørensen, Daniel Mosgaard and Tankisi, Hatice},
  title = {Reliability of MScanFit Motor Unit Number Estimation in the Trapezius Muscle},
  journal = {Muscle & Nerve},
  year = {2025},
  doi = {10.1002/mus.28303}
}
Kostenlos starten →