EUVIMEDEUVIMED entdecken
Crossref

Recording sensory nerve action potential using different electrode types

Sanjeev D Nandedkar iD, Paul E Barkhaus

Muscle & Nerve · 2023

DOI: 10.1002/mus.27916 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Muscle & Nerve, 2023. Verfasst von Sanjeev D Nandedkar und 1 weiteren Autor:innen. Beteiligte Einrichtung: Natus Medical Inc, Middleton, Wisconsin, USA. Hauptthemen laut MEDLINE: Evoked Potentials, Neural Conduction. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 8 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction/Aims Switching between different types of electrodes during motor and sensory nerve conduction studies adds time to a study. We investigated the use of disposable disc electrodes (DDE) used for motor nerve conduction studies to record the antidromic sensory nerve action potential (SNAP) in median, ulnar and radial sensory nerve conduction studies.

Methods

The SNAP was recorded using four different electrode types: reusable ring, reusable bar, disposable ring, and DDE in a random rotating order. Studies were performed in healthy subjects. Other than being an adult with no history of neuromuscular disease, there were no exclusion criteria.

Results

We studied 20 subjects (11 females, 9 males; age 41.1 ± 15.7 y). The SNAP waveforms recorded by all four electrode types were similar. There was no statistically significant difference in the onset latency, peak latency (PL), negative peak amplitude (NPA), peak to peak amplitude, or conduction velocity. In individual nerve recordings, the absolute PL difference between reusable ring electrodes (our current standard) and DDE was less than 0.2 ms in 58 of 60 (97%) nerves. The mean absolute NPA difference was 3.1 μV (standard deviation = 2.85 μV). Recordings with NPA difference >5 μV also had high NPA and/or had large artifacts.

Discussion

DDE may be used for performing motor and sensory nerve conduction studies. This can reduce the time required for electrodiagnostic testing.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Evoked PotentialsNeural ConductionAdultFemaleHumansMaleMiddle AgedAction PotentialsElectrodesMedian NerveReaction TimeUlnar Nerve

Stichwörter der Autor:innen

amplitudeelectrodeselectrodiagnosissensory nerve action potentialsensory nerve conduction studytechnique

Autor:innen und Institutionen

  • Natus Medical Inc, Middleton, Wisconsin, USA.
  • Departments of Neurology, Medical College of Wisconsin, Milwaukee, Wisconsin, USA.

ORCID-Profile: Sanjeev D Nandedkar

Zitieren

APA

Nandedkar, S. D., & Barkhaus, P. E. (2023). Recording sensory nerve action potential using different electrode types. Muscle & Nerve. https://doi.org/10.1002/mus.27916

Vancouver

Nandedkar SD, Barkhaus PE. Recording sensory nerve action potential using different electrode types. Muscle & Nerve. 2023. doi:10.1002/mus.27916

BibTeX
@article{Nandedkar2023,
  author = {Nandedkar, Sanjeev D and Barkhaus, Paul E},
  title = {Recording sensory nerve action potential using different electrode types},
  journal = {Muscle & Nerve},
  year = {2023},
  doi = {10.1002/mus.27916}
}
Kostenlos starten →