Ultimate Intrinsic SNR Depends on B 0 Orientation

Yuting Wang iD, Markus W May iD, Marcel Gratz, Mark E Ladd, Stephan Orzada

Magnetic Resonance in Medicine · 2027-01 · 97 (1) · 340-350

DOI: 10.1002/mrm.70577 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, online erschienen am 25.08.2026. Verfasst von Yuting Wang und 4 weiteren Autor:innen. Beteiligte Einrichtung: Medical Physics in Radiology, German Cancer Research Center (DKFZ), Heidelberg, Germany. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Image Processing, Computer-Assisted. Die Arbeit nennt 33 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: HORIZON EUROPE Framework Programme.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

This work aims to investigate the effect of orientation on the uiSNR in MRI.

Methods

Electromagnetic simulations were performed on both realistic body models and simplified geometric models. Comparisons between (1) opposite directions along the head‐foot (HF) axis, corresponding to head‐first and feet‐first patient orientations, at various field strengths, and (2) along the HF, anterior–posterior (AP), and left–right (LR) directions at 0.55 T, corresponding to closed‐bore (HF) and open (AP/LR) MR systems. The voxel‐wise ratio of uiSNR between different orientations was calculated to characterize spatial variations throughout the models.

Results

For closed‐bore MRI systems, flipping the patient orientation results in a spatial redistribution of uiSNR, and the magnitude of this redistribution increases with field strength. At 0.55 T, open MR systems with either AP‐ or LR‐oriented yield higher uiSNR than closed‐bore systems in the center of cylindrical‐shaped structures. In realistic body models, LR‐oriented yields higher uiSNR than AP‐oriented in most regions at 0.55 T, with both orientations outperforming HF‐oriented in the center of the body.

Conclusion

The uiSNR distribution varies with orientation in realistic body models. At around 0.55 T, the flexibility in orientation enabled by different magnet configurations may serve as an additional degree of freedom for improving imaging performance, as uiSNR is strongly influenced by the orientation of the main field.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Magnetic Resonance ImagingImage Processing, Computer-AssistedHumansComputer SimulationSignal-To-Noise RatioPhantoms, ImagingHeadAlgorithms

Stichwörter der Autor:innen

B0 orientationMRIin silicopatient orientationuiSNR

Autor:innen und Institutionen

  • Medical Physics in Radiology, German Cancer Research Center (DKFZ), Heidelberg, Germany.
  • Faculty of Physics and Astronomy, Heidelberg University, Heidelberg, Germany.
  • High-Field and Hybrid MR Imaging, University Hospital Essen, Essen, Germany.
  • Erwin L. Hahn Institute for MR Imaging, University of Duisburg-Essen, Essen, Germany.
  • Faculty of Medicine, Heidelberg University, Heidelberg, Germany.

ORCID-Profile: Yuting Wang, Markus W May

Förderung

  • HORIZON EUROPE Framework Programme

Zitieren

APA

Wang, Y., May, M. W., Gratz, M., Ladd, M. E., & Orzada, S. (2027). Ultimate Intrinsic SNR Depends on B 0 Orientation. Magnetic Resonance in Medicine, 97(1), 340-350. https://doi.org/10.1002/mrm.70577

Vancouver

Wang Y, May MW, Gratz M, Ladd ME, Orzada S. Ultimate Intrinsic SNR Depends on B 0 Orientation. Magnetic Resonance in Medicine. 2027;97(1):340-350. doi:10.1002/mrm.70577

BibTeX
@article{Wang2027,
  author = {Wang, Yuting and May, Markus W and Gratz, Marcel and Ladd, Mark E and Orzada, Stephan},
  title = {Ultimate Intrinsic SNR Depends on B 0 Orientation},
  journal = {Magnetic Resonance in Medicine},
  year = {2027},
  volume = {97},
  number = {1},
  pages = {340--350},
  doi = {10.1002/mrm.70577}
}
Kostenlos starten →