A Flexible Graphite–Silicone Pad for Improved Fat Suppression in Cervical Spine MRI at 5 T via Passive B 0 Shimming
Magnetic Resonance in Medicine · 2026
DOI: 10.1002/mrm.70535 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2026. Verfasst von Yihan Yang und 4 weiteren Autor:innen. Beteiligte Einrichtung: School of Biomedical Engineering & State Key Laboratory of Advanced Medical Materials and Devices, ShanghaiTech University, Shanghai, China. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Graphite, Cervical Vertebrae, Adipose Tissue. Die Arbeit nennt 26 Literaturangaben. Gefördert von: Shanghai Pujiang Program, The Explorers Program of Shanghai-Basic Research Funding, National Natural Science Foundation of China.
Abstract (Originalsprache)
Purpose
Susceptibility‐induced inhomogeneity at the posterior neck degrades spectral fat suppression in cervical spine MRI at 5 T and above. We describe a flexible graphite–silicone pad for passive shimming and assess its effect on field homogeneity and fat‐suppressed imaging alongside the scanner's active shim.
Methods
Pads were fabricated by dispersing high‐purity graphite powder in a biocompatible silicone elastomer and casting the mixture in a 3D‐printed mold shaped to the posterior neck and head–neck receive coil. Five graphite‐to‐silicone mass ratios (1.5:100 through 8:100) were screened in one volunteer on a 5 T scanner using dual‐echo mapping and fat‐suppressed ‐ and ‐weighted FSE. The selected ratio (4.5:100) was evaluated in five healthy volunteers under volume‐based active shimming, with and without the pad in randomized order.
Results
In the screening volunteer under Volume shimming, the 4.5:100 pad reduced the posterior‐neck field from 0.740 to 0.458 ppm, a 38% reduction at matched shim mode. Across the five validation subjects, decreased in every case under Volume shimming, from to ppm (individual reductions 41%–55%). Residual hyperintense fat seen with the standard pad was no longer present on sagittal ‐ or ‐weighted images, and the improvement extended to the supraclavicular region on coronal ‐weighted images.
Conclusion
A flexible graphite–silicone pad complements active shimming and improves spectral fat suppression in cervical spine MRI at 5 T. Evaluation in patients and on EPI‐based sequences remains to be done.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- School of Biomedical Engineering & State Key Laboratory of Advanced Medical Materials and Devices, ShanghaiTech University, Shanghai, China.
- United Imaging Healthcare, Shanghai, China.
- Shanghai Clinical Research and Trial Center, Shanghai, China.
ORCID-Profile: Yihan Yang, Haotian Hong, Zhihua Ren
Förderung
- Shanghai Pujiang Program
- The Explorers Program of Shanghai-Basic Research Funding
- National Natural Science Foundation of China
Zitieren
APA
Yang, Y., Liu, Z., Hong, H., Zhang, W., & Ren, Z. (2026). A Flexible Graphite–Silicone Pad for Improved Fat Suppression in Cervical Spine MRI at 5 T via Passive B 0 Shimming. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.70535
Vancouver
Yang Y, Liu Z, Hong H, Zhang W, Ren Z. A Flexible Graphite–Silicone Pad for Improved Fat Suppression in Cervical Spine MRI at 5 T via Passive B 0 Shimming. Magnetic Resonance in Medicine. 2026. doi:10.1002/mrm.70535
BibTeX
@article{Yang2026,
author = {Yang, Yihan and Liu, Ziyu and Hong, Haotian and Zhang, Weijun and Ren, Zhihua},
title = {A Flexible Graphite–Silicone Pad for Improved Fat Suppression in Cervical Spine MRI at 5 T via Passive B 0 Shimming},
journal = {Magnetic Resonance in Medicine},
year = {2026},
doi = {10.1002/mrm.70535}
}
EUVIMED