EUVIMEDEUVIMED entdecken
Open Access

Whole‐Body Oxygen‐Enhanced MRI of Tumor‐Bearing Mice Under Hyperbaric Oxygenation

Conner S Ubert iD, Maciej M Kmiec, Paula S Lara Mejia, Alireza Kheirollah, Ryan C O'Connell, Sergey V Petryakov, Philip E Schaner, Periannan Kuppusamy iD

Magnetic Resonance in Medicine · 2026-10 · 96 (4) · 1860-1871

DOI: 10.1002/mrm.70458 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, online erschienen am 11.06.2026. Verfasst von Conner S Ubert und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Geisel School of Medicine, Dartmouth College, Hanover, New Hampshire, USA. Hauptthemen laut MEDLINE: Hyperbaric Oxygenation, Oxygen, Magnetic Resonance Imaging, Whole Body Imaging. Die Arbeit nennt 41 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NCI NIH HHS, NIH HHS, NIBIB NIH HHS.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Oxygen‐enhanced MRI (OE‐MRI) enables noninvasive assessment of tissue oxygenation by exploiting the paramagnetic effect of molecular oxygen on longitudinal relaxation. The purpose of this study was to evaluate whether whole‐body oxygen‐enhanced MRI (OE‐MRI) can be performed under hyperbaric oxygen (HBO) conditions and to assess its utility for detecting oxygenation changes in a small‐animal tumor model using a clinical 3 T scanner.

Methods

A fully MRI‐compatible HBO chamber was built for imaging mice and installed in a 3 T clinical scanner. SCC7‐tumor‐bearing C3H mice underwent whole‐body OE‐MRI with variable flip‐angle T 1 mapping. Longitudinal relaxation‐rates ( R 1 ) and Δ R 1 were measured while the animals breathed 21% O 2 at 1 ATA, 21% O 2 at 2 ATA, 100% O 2 at 1 ATA, and 100% O 2 at 2 ATA. Δ R 1 was quantified in tumor, brain, and heart regions.

Results

High‐quality OE‐MRI was obtained under all hyperbaric conditions. Normal tissues (brain, heart) showed uniform, pressure‐dependent increases in Δ R 1 with higher inspired oxygen fractions. Tumors displayed pronounced spatial heterogeneity: many voxels responded robustly, while others showed little or no Δ R 1 change even at 2 ATA 100% O 2 . The proportion of oxygen‐responsive tumor voxels rose with increasing pressure, yet non‐responsive regions persisted.

Conclusion

Whole‐body pre‐clinical OE‐MRI on a clinical 3 T system is feasible and sensitive under hyperbaric oxygenation, capturing both tissue‐level pressure effects and heterogeneous tumor oxygen responses. This non‐invasive HBO‐OE‐MRI platform provides a valuable pre‐clinical tool for assessing oxygen delivery and identifying oxygen‐refractory tumor zones, supporting oxygen‐guided therapeutic strategies.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Hyperbaric OxygenationOxygenMagnetic Resonance ImagingWhole Body ImagingAnimalsMiceMice, Inbred C3HCell Line, TumorBrain

Stichwörter der Autor:innen

EPRHBOMRIOE‐MRImouseoximetrypO2

Substanzen

Oxygen

Autor:innen und Institutionen

  • Department of Radiology, Geisel School of Medicine, Dartmouth College, Hanover, New Hampshire, USA.
  • Thayer School of Engineering, Dartmouth College, Hanover, New Hampshire, USA.
  • Department of Radiation Oncology and Applied Sciences, Geisel School of Medicine, Dartmouth College, Hanover, New Hampshire, USA.
  • Department of Surgery, Geisel School of Medicine, Dartmouth College, Hanover, New Hampshire, USA.

ORCID-Profile: Conner S Ubert, Periannan Kuppusamy

Förderung

  • NCI NIH HHS
  • NIH HHS
  • NIBIB NIH HHS
  • National Cancer Institute
  • National Institutes of Health

Zitieren

APA

Ubert, C. S., Kmiec, M. M., Lara Mejia, P. S., Kheirollah, A., O'Connell, R. C., Petryakov, S. V., Schaner, P. E., & Kuppusamy, P. (2026). Whole‐Body Oxygen‐Enhanced MRI of Tumor‐Bearing Mice Under Hyperbaric Oxygenation. Magnetic Resonance in Medicine, 96(4), 1860-1871. https://doi.org/10.1002/mrm.70458

Vancouver

Ubert CS, Kmiec MM, Lara Mejia PS, Kheirollah A, O'Connell RC, Petryakov SV, et al. Whole‐Body Oxygen‐Enhanced MRI of Tumor‐Bearing Mice Under Hyperbaric Oxygenation. Magnetic Resonance in Medicine. 2026;96(4):1860-1871. doi:10.1002/mrm.70458

BibTeX
@article{Ubert2026,
  author = {Ubert, Conner S and Kmiec, Maciej M and Lara Mejia, Paula S and Kheirollah, Alireza and O'Connell, Ryan C and Petryakov, Sergey V and Schaner, Philip E and Kuppusamy, Periannan},
  title = {Whole‐Body Oxygen‐Enhanced MRI of Tumor‐Bearing Mice Under Hyperbaric Oxygenation},
  journal = {Magnetic Resonance in Medicine},
  year = {2026},
  volume = {96},
  number = {4},
  pages = {1860--1871},
  doi = {10.1002/mrm.70458}
}
Kostenlos starten →