Reducing Motion Artifact in High Resolution 7 T MRI Using the Magnetic Resonance Minimal Motion (“MR‐MinMo”) Head Stabilization Device
Magnetic Resonance in Medicine · 2026-09 · 96 (3) · 1486-1501
DOI: 10.1002/mrm.70424 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, online erschienen am 02.06.2026. Verfasst von Jyoti Mangal und 10 weiteren Autor:innen. Beteiligte Einrichtung: Research Department of Imaging Physics and Engineering, School of Biomedical Engineering and Imaging Sciences, King's College London, London, UK. Hauptthemen laut MEDLINE: Artifacts, Magnetic Resonance Imaging, Brain, Head. Die Arbeit nennt 55 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: EPSRC smart medical imaging CDT studentship, Wellcome/EPSRC Centre for Medical Engineering, National Institute for Health Research (NIHR) Biomedical Research Centre based at Guy's and St Thomas' NHS Foundation Trust and King's College London.
Abstract (Originalsprache)
Purpose
To evaluate the effectiveness of the MR‐MinMo head stabilization device in mitigating motion artifacts during long‐duration, high‐resolution 7 T MRI scans, with and without retrospective motion correction Methods The MR‐MinMo was tested on seven pediatric and 12 adult healthy volunteers using 0.6 mm isotropic, 3D Multi‐Echo Gradient Echo (ME‐GRE) scans. A ∼10‐min ME‐GRE scan (linear sampling acceleration factor 2 × 2) and a ∼20‐min scan (DISORDER sampling acceleration factor 1.4 × 1.4) were obtained. Both scans were acquired with and without the MR‐MinMo in a 2 × 2 factorial study design. Qualitative and quantitative assessment of image quality of the first echo image was obtained via visual inspection and the normalized gradient squared (NGS) metric, respectively. A repeated measures ANOVA was used to determine individual and combined effects of the device and DISORDER images with retrospective motion correction on NGS values, with age as a binary factor (pediatric/adult). T2* maps were generated and the standard deviation of white matter R2*(=1/T2*) values assessed across conditions.
Results
The MR‐MinMo significantly reduced motion artifacts both visually and in terms of NGS scores, particularly in pediatric volunteers. There was a significant interaction between MR‐MinMo and DISORDER motion correction, suggesting that MR‐MinMo improved retrospective motion correction. T2* maps demonstrated improved visual appearance and reduced WM variance with the MR‐MinMo.
Conclusion
The MR‐MinMo can improve image quality via motion reduction in high‐resolution 7 T scans. By keeping motion within a correctable regime, the device can also improve the performance of retrospective motion correction methods.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Research Department of Imaging Physics and Engineering, School of Biomedical Engineering and Imaging Sciences, King's College London, London, UK.
- Division of Psychology and Language Sciences, University College London, London, UK.
- Wellcome Centre for Human Neuroimaging, University College London, London, UK.
ORCID-Profile: Jyoti Mangal, Yannick Brackenier, Shaihan Malik, Martina F Callaghan
Förderung
- EPSRC smart medical imaging CDT studentship
- Wellcome/EPSRC Centre for Medical Engineering
- National Institute for Health Research (NIHR) Biomedical Research Centre based at Guy's and St Thomas' NHS Foundation Trust and King's College London
- National Institute for Health and Care Research
Zitieren
APA
Mangal, J., Richardson, S., Brackenier, Y., Gardner, M., Di Cio, P., Casella, C., Malik, S., Hajnal, J., Callaghan, M. F., Dick, F., & Carmichael, D. W. (2026). Reducing Motion Artifact in High Resolution 7 T MRI Using the Magnetic Resonance Minimal Motion (“MR‐MinMo”) Head Stabilization Device. Magnetic Resonance in Medicine, 96(3), 1486-1501. https://doi.org/10.1002/mrm.70424
Vancouver
Mangal J, Richardson S, Brackenier Y, Gardner M, Di Cio P, Casella C, et al. Reducing Motion Artifact in High Resolution 7 T MRI Using the Magnetic Resonance Minimal Motion (“MR‐MinMo”) Head Stabilization Device. Magnetic Resonance in Medicine. 2026;96(3):1486-1501. doi:10.1002/mrm.70424
BibTeX
@article{Mangal2026,
author = {Mangal, Jyoti and Richardson, Simon and Brackenier, Yannick and Gardner, Matthew and Di Cio, Pierluigi and Casella, Chiara and Malik, Shaihan and Hajnal, Jo and Callaghan, Martina F and Dick, Fred and Carmichael, David W},
title = {Reducing Motion Artifact in High Resolution 7 T MRI Using the Magnetic Resonance Minimal Motion (“MR‐MinMo”) Head Stabilization Device},
journal = {Magnetic Resonance in Medicine},
year = {2026},
volume = {96},
number = {3},
pages = {1486--1501},
doi = {10.1002/mrm.70424}
}
EUVIMED