Dependence of Magnetic Resonance Elastography‐Derived Stiffness Values on Selected Imaging Parameters
Magnetic Resonance in Medicine · 2026-09 · 96 (3) · 1083-1096
DOI: 10.1002/mrm.70391 ↗Kurz gefasst
Beobachtungsstudie in der Zeitschrift Magnetic Resonance in Medicine, online erschienen am 22.04.2026. Verfasst von Łukasz Hańczyk und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Independent Public Health Care Center of the Ministry of Interior and Administration in Rzeszów, Rzeszów, Poland. Hauptthemen laut MEDLINE: Elasticity Imaging Techniques, Liver, Magnetic Resonance Imaging. Die Arbeit nennt 23 Literaturangaben. Gefördert von: Uniwersytet Rzeszowski, Narodowa Agencja Wymiany Akademickiej.
Abstract (Originalsprache)
Purpose
To evaluate the influence of driver power and slice thickness on magnetic resonance elastography (MRE)‐derived liver stiffness in human subjects and a tissue‐mimicking phantom.
Methods
A prospective single‐center observational study was performed involving 34 adult participants (20 males; median age: 29.0 ± 29.3 years) and a standard phantom. MRE was conducted on a 1.5T scanner using a 2D gradient‐recalled echo‐based sequence. Human data were acquired at driver powers of 25%–100% and slice thicknesses of 5 and 10 mm. The phantom was scanned with driver powers of 10%–100% and slice thicknesses from 2 to 10 mm. Two region‐of‐interest (ROI) approaches were applied: (A) fixed‐location ROIs and (B) largest measurable ROIs. Stiffness maps were generated using the MMDI inversion algorithm. Statistical analysis included repeated measures ANOVA, paired t‐tests or Wilcoxon signed‐rank tests, and intraclass correlation coefficients (ICCs).
Results
Measured whole‐liver stiffness varied modestly over the driver power range, with the effect reaching statistical significance in Method B analysis ( p = 0.037). Measurements of individual slices varied more significantly with driver power (Method A: p = 0.031; Method B: p < 0.001). Slice thickness had a significant effect on individual slice measurements (Method B: p = 0.007) but not on whole‐liver measurements. In phantom studies, both parameters significantly influenced stiffness ( p < 0.001 to p = 0.017). ICCs indicated good to excellent repeatability in vivo (0.79–0.96) and moderate to good repeatability in the phantom (0.62–0.89).
Conclusion
The results demonstrate that measured MRE‐derived liver stiffness can be influenced by driver power and slice thickness, underscoring the need for standardization of these parameters in both clinical practice and research.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiology, Independent Public Health Care Center of the Ministry of Interior and Administration in Rzeszów, Rzeszów, Poland.
- Laboratory of Innovative Medical Imaging Methods, Faculty of Health Sciences and Psychology, Medical College, University of Rzeszów, Rzeszów, Poland.
- Institute of Physics, Faculty of Exact and Technical Sciences, University of Rzeszów, Rzeszów, Poland.
- Department of Physics and Medical Engineering, Faculty of Mathematics and Applied Physics, Rzeszów University of Technology, Rzeszów, Poland.
Förderung
- Uniwersytet Rzeszowski
- Narodowa Agencja Wymiany Akademickiej
Zitieren
APA
Hańczyk, Ł., Obrzut, M., Cholewa, M., & Atamaniuk, V. (2026). Dependence of Magnetic Resonance Elastography‐Derived Stiffness Values on Selected Imaging Parameters. Magnetic Resonance in Medicine, 96(3), 1083-1096. https://doi.org/10.1002/mrm.70391
Vancouver
Hańczyk Ł, Obrzut M, Cholewa M, Atamaniuk V. Dependence of Magnetic Resonance Elastography‐Derived Stiffness Values on Selected Imaging Parameters. Magnetic Resonance in Medicine. 2026;96(3):1083-1096. doi:10.1002/mrm.70391
BibTeX
@article{Haczyk2026,
author = {Hańczyk, Łukasz and Obrzut, Marzanna and Cholewa, Marian and Atamaniuk, Vitaliy},
title = {Dependence of Magnetic Resonance Elastography‐Derived Stiffness Values on Selected Imaging Parameters},
journal = {Magnetic Resonance in Medicine},
year = {2026},
volume = {96},
number = {3},
pages = {1083--1096},
doi = {10.1002/mrm.70391}
}
EUVIMED