3D MR elastography at 0. 55 T : Concomitant field effects and feasibility
Magnetic Resonance in Medicine · 2025
DOI: 10.1002/mrm.30377 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2025. Verfasst von Omar Isam Darwish und 3 weiteren Autor:innen. Beteiligte Einrichtung: Research Department of Imaging Physics and Engineering, School of Biomedical Engineering and Imaging Sciences, King's College London, London, UK. Hauptthemen laut MEDLINE: Elasticity Imaging Techniques, Imaging, Three-Dimensional, Liver, Magnetic Resonance Imaging. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 41 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Wellcome Trust, CRUK, Wellcome/EPSRC Centre for Medical Engineering.
Abstract (Originalsprache)
Purpose
To demonstrate the feasibility of hepatic 3D MR elastography (MRE) at 0.55 T in healthy volunteers using Hadamard encoding and to study the effects of concomitant fields in the domain of MRE in general.
Methods
Concomitant field effects in MRE are assessed using a Taylor series expansion and an encoding scheme is proposed to study the corresponding effects on 3D MRE at 0.55 T in numerical simulations and in phantom experiments. In addition, five healthy volunteers were enrolled and scanned at 60 Hz mechanical excitation with a Hadamard‐encoded 3D MRE sequence at 0.55 T and were also scanned with a reference 3D MRE sequence at 3 T for comparison. The retrieved biomechanical parameters were the magnitude of the complex shear modulus (| G *|), the shear wave speed (Cs), and the loss modulus ( G ″). Comparison of apparent SNR between 3 T and 0.55 T was performed.
Results
Theoretical analysis, numerical simulations and phantom experiments demonstrated that the effects of concomitant fields in 3D MRE at 0.55 T are negligible. In the healthy volunteer experiments, the mean values of | G *|, Cs, and G″ in the liver were 2.1 ± 0.3 kPa, 1.5 ± 0.1 m/s, and 0.8 ± 0.1 kPa at 0.55 T, respectively, and 2.0 ± 0.2 kPa, 1.5 ± 0.1 m/s, and 0.9 ± 0.1 kPa at 3 T, respectively. Bland–Altman analysis demonstrated good agreement between the biomechanical parameters retrieved at 0.55 T and 3 T. A 2.1‐fold relative apparent SNR decrease was observed in 3D MRE at 0.55 T in comparison with 3 T.
Conclusion
Hepatic 3D MRE is feasible at 0.55 T, showing promising initial results in healthy volunteers.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Research Department of Imaging Physics and Engineering, School of Biomedical Engineering and Imaging Sciences, King's College London, London, UK.
- MR Predevelopment, Siemens Healthineers AG, Erlangen, Germany.
- INSERM U1148, LVTS, University Paris Diderot, Paris, France.
ORCID-Profile: Omar Isam Darwish
Förderung
- Wellcome Trust
- CRUK
- Wellcome/EPSRC Centre for Medical Engineering
- Siemens Healthineers
- Cancer Research UK
- Wellcome EPSRC Centre for Medical Engineering
Zitieren
APA
Darwish, O. I., Di Cio, P., Sinkus, R., & Neji, R. (2025). 3D MR elastography at 0. 55 T : Concomitant field effects and feasibility. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30377
Vancouver
Darwish OI, Di Cio P, Sinkus R, Neji R. 3D MR elastography at 0. 55 T : Concomitant field effects and feasibility. Magnetic Resonance in Medicine. 2025. doi:10.1002/mrm.30377
BibTeX
@article{Darwish2025,
author = {Darwish, Omar Isam and Di Cio, Pierluigi and Sinkus, Ralph and Neji, Radhouene},
title = {3D MR elastography at 0. 55 T : Concomitant field effects and feasibility},
journal = {Magnetic Resonance in Medicine},
year = {2025},
doi = {10.1002/mrm.30377}
}
EUVIMED