Correcting 129 Xe Gas Exchange MRI for Incidental Gas‐Phase Excitation—Comparing Approaches, and Identifying Acceptable Thresholds for Reliable Quantification
Magnetic Resonance in Medicine · 2026-08 · 96 (2) · 787-798
DOI: 10.1002/mrm.70390 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, online erschienen am 20.04.2026. Verfasst von Suphachart Leewiwatwong und 5 weiteren Autor:innen. Beteiligte Einrichtung: Biomedical Engineering, Duke University, Durham, North Carolina, USA. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Xenon Isotopes, Image Processing, Computer-Assisted. Die Arbeit nennt 39 Literaturangaben. Gefördert von: NIH/NHLBI.
Abstract (Originalsprache)
Purpose
Incidental excitation of the gas‐phase resonance during 129 Xe gas exchange MRI introduces contamination that can bias membrane uptake and red blood cell (RBC) transfer metrics. Here, we defined the level of gas‐phase contamination that is acceptable, and compared the dual‐echo (Hahn) and spectroscopy‐informed (Willmering) frameworks to correct for it.
Methods
Nine healthy volunteers underwent interleaved 3D radial dual‐echo Dixon‐based 129 Xe MRI and spectroscopy. The dissolved‐phase RF pulse was shortened from its optimal value to increase incidental gas contamination. The Hahn and Willmering corrections were implemented and refined using a grid search over a contamination‐dominated ROI to determine any platform‐specific scale and phase adjustments. The effects of contamination phase and scale were tested by synthetically reintroducing gas signal at phase offsets that maximized defect‐ and high intensity‐like artifacts. Membrane/gas and RBC/gas defect and high percentages were evaluated against thresholds derived from a healthy reference cohort to determine acceptable gas contamination.
Results
Images acquired with the shortened pulse contained 24% ± 6% gas contamination. After applying empirically calibrated phase shifts (Hahn≈230°, Willmering≈49°) and no scale adjustment (optimal value≈1), both correction methods increased RBC SNR by ∼9% and reduced background noise. Phase‐ and scale‐sweep experiments indicated that gas contamination becomes unacceptable once it exceeds ∼9% ± 3%, at which point membrane and RBC defect or high percentages surpass their thresholds.
Conclusions
Gas‐phase contamination beyond ∼9% requires correction. After a one‐time empirical phase calibration, this can be equally well achieved by both the Hahn and Willmering methods.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Biomedical Engineering, Duke University, Durham, North Carolina, USA.
- Division of Pulmonary Medicine, Cincinnati Children's Hospital Medical Center, Cincinnati, Ohio, USA.
- Department of Pediatrics, University of Cincinnati, Cincinnati, Ohio, USA.
- Department of Radiology, University of Iowa, Iowa City, Iowa, USA.
- Department of Radiology & Medical Imaging, University of Virginia School of Medicine, Charlottesville, Virginia, USA.
- Department of Radiology, Duke University, Durham, North Carolina, USA.
ORCID-Profile: Suphachart Leewiwatwong, Matthew Willmering, Andrew D Hahn
Förderung
- NIH/NHLBI
Zitieren
APA
Leewiwatwong, S., Willmering, M., Hahn, A. D., Mugler, J. P., Mummy, D., & Driehuys, B. (2026). Correcting 129 Xe Gas Exchange MRI for Incidental Gas‐Phase Excitation—Comparing Approaches, and Identifying Acceptable Thresholds for Reliable Quantification. Magnetic Resonance in Medicine, 96(2), 787-798. https://doi.org/10.1002/mrm.70390
Vancouver
Leewiwatwong S, Willmering M, Hahn AD, Mugler JP, Mummy D, Driehuys B. Correcting 129 Xe Gas Exchange MRI for Incidental Gas‐Phase Excitation—Comparing Approaches, and Identifying Acceptable Thresholds for Reliable Quantification. Magnetic Resonance in Medicine. 2026;96(2):787-798. doi:10.1002/mrm.70390
BibTeX
@article{Leewiwatwong2026,
author = {Leewiwatwong, Suphachart and Willmering, Matthew and Hahn, Andrew D and Mugler, John P and Mummy, David and Driehuys, Bastiaan},
title = {Correcting 129 Xe Gas Exchange MRI for Incidental Gas‐Phase Excitation—Comparing Approaches, and Identifying Acceptable Thresholds for Reliable Quantification},
journal = {Magnetic Resonance in Medicine},
year = {2026},
volume = {96},
number = {2},
pages = {787--798},
doi = {10.1002/mrm.70390}
}
EUVIMED