EUVIMEDEUVIMED entdecken
Crossref

Noncontrast MR Imaging of Blood–Brain Barrier Permeability: Halving Scan Time With Velocity‐Selective Preparation for Vessel Isolation

Yizhe Hu, Yi-Cheng Hsu, Dan Wu iD, Xihai Zhao, Qin Qin iD, Hanzhang Lu, Li Zhao, Zixuan Lin iD

Magnetic Resonance in Medicine · 2026-08 · 96 (2) · 568-579

DOI: 10.1002/mrm.70349 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, online erschienen am 24.03.2026. Verfasst von Yizhe Hu und 7 weiteren Autor:innen. Beteiligte Einrichtung: Key Laboratory for Biomedical Engineering of the Ministry of Education, Department of Biomedical Engineering, College of Biomedical Engineering & Ins…. Hauptthemen laut MEDLINE: Blood-Brain Barrier, Magnetic Resonance Imaging, Image Processing, Computer-Assisted. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 67 Literaturangaben. Gefördert von: National Natural Science Foundation of China, Fundamental Research Funds for the Central Universities.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

This study aims to develop and evaluate a noncontrast MRI method, Water‐Extraction‐with‐Fourier‐transform‐velocity‐selective‐Arterial‐Spin‐Tagging (WEFAST), for accelerated and robust measurement of blood–brain barrier (BBB) permeability to water.

Methods

The proposed WEFAST sequence incorporates Fourier‐Transform Velocity‐Selective (FTVS) preparation into the Water‐Extraction‐with‐Phase‐Contrast‐Arterial‐Spin‐Tagging (WEPCAST) MRI framework, enabling single‐shot acquisition without the need for double‐subtraction measurement. The imaging parameters of WEFAST were optimized for maximizing spatial and temporal SNR. Its reproducibility and evaluation against the established WEPCAST technique were performed.

Results

Optimal scan parameters were determined to be velocity cut‐off = 2 cm/s and FTVS flip angle = 90°, and the spatial SNR reached 10.42 ± 1.30 which is significantly higher than that of other imaging protocols (all p < 0.01). The scan time of WEFAST was significantly reduced (WEFAST vs. WEPCAST: 3 min 50 s vs. 6 min 54 s). The WEFAST sequence preserved robust reproducibility (intrasession CoV: extraction fraction ( E ) = 3.11% ± 0.67%, permeability surface area product (PS) = 11.08% ± 1.29%; intersession CoV: E = 3.17% ± 0.76%, PS = 12.33% ± 2.04%) and achieved good agreement with the WEPCAST in quantitative measurements ( R = 0.59, p < 0.001 for E ; R = 0.74, p < 0.001 for PS). Multidelay experiments further confirmed reliable quantification of venous water extraction signals in multiple venous structures such as the transverse sinus and straight sinus.

Conclusion

The proposed WEFAST sequence achieved significant acceleration in noncontrast BBB MR imaging by providing rapid and reliable quantification of BBB water permeability, with the potential for clinical applications.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Blood-Brain BarrierMagnetic Resonance ImagingImage Processing, Computer-AssistedReproducibility of ResultsHumansAlgorithmsSignal-To-Noise RatioFourier AnalysisSpin LabelsPhantoms, ImagingPermeabilityWater

Stichwörter der Autor:innen

BBB permeabilityFourier‐transform velocity‐selectivearterial spin labelingblood–brain barrierwater‐extraction‐with‐phase‐contrast‐arterial‐spin‐tagging

Substanzen

Spin LabelsWater

Autor:innen und Institutionen

  • Key Laboratory for Biomedical Engineering of the Ministry of Education, Department of Biomedical Engineering, College of Biomedical Engineering & Instrument Science, Zhejiang University, Hangzhou, China.
  • MR Collaboration, Siemens Healthineers Ltd., Shanghai, China.
  • Center for Biomedical Imaging Research, School of Biomedical Engineering, Tsinghua University, Beijing, China.
  • Department of Radiology, University of Washington, Seattle, Washington, USA.
  • The Russell H. Morgan Department of Radiology & Radiological Science, Johns Hopkins University School of Medicine, Baltimore, Maryland, USA.

ORCID-Profile: Dan Wu, Qin Qin, Zixuan Lin

Förderung

  • National Natural Science Foundation of China
  • Fundamental Research Funds for the Central Universities

Zitieren

APA

Hu, Y., Hsu, Y. C., Wu, D., Zhao, X., Qin, Q., Lu, H., Zhao, L., & Lin, Z. (2026). Noncontrast MR Imaging of Blood–Brain Barrier Permeability: Halving Scan Time With Velocity‐Selective Preparation for Vessel Isolation. Magnetic Resonance in Medicine, 96(2), 568-579. https://doi.org/10.1002/mrm.70349

Vancouver

Hu Y, Hsu YC, Wu D, Zhao X, Qin Q, Lu H, et al. Noncontrast MR Imaging of Blood–Brain Barrier Permeability: Halving Scan Time With Velocity‐Selective Preparation for Vessel Isolation. Magnetic Resonance in Medicine. 2026;96(2):568-579. doi:10.1002/mrm.70349

BibTeX
@article{Hu2026,
  author = {Hu, Yizhe and Hsu, Yi-Cheng and Wu, Dan and Zhao, Xihai and Qin, Qin and Lu, Hanzhang and Zhao, Li and Lin, Zixuan},
  title = {Noncontrast MR Imaging of Blood–Brain Barrier Permeability: Halving Scan Time With Velocity‐Selective Preparation for Vessel Isolation},
  journal = {Magnetic Resonance in Medicine},
  year = {2026},
  volume = {96},
  number = {2},
  pages = {568--579},
  doi = {10.1002/mrm.70349}
}
Kostenlos starten →