Relaxivity of Gadobutrol and Gadoteric Acid in Cerebrospinal Fluid at 3 T

Sofia Behndig iD, Anders Garpebring iD, Daniel Dahlgren Lindström iD, Jan Malm, Anders Wåhlin, Anders Eklund

Magnetic Resonance in Medicine · 2026

DOI: 10.1002/mrm.70284 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2026. Verfasst von Sofia Behndig und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Diagnostics and Intervention, Biomedical Engineering and Radiation Physics, Umeå University, Umeå, Sweden. Hauptthemen laut MEDLINE: Organometallic Compounds, Contrast Media, Magnetic Resonance Imaging, Cerebrospinal Fluid. Die Arbeit nennt 38 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Vetenskapsrådet, Stiftelsen för Strategisk Forskning.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

The aim was to estimate T1 relaxivity of gadobutrol and gadoteric acid in cerebrospinal fluid (CSF) at 3T, to support research on CSF‐flow and the glymphatic system in humans utilizing T1 mapping after intrathecal injection.

Methods

Using a phantom, relaxivity was estimated for gadobutrol and gadoteric acid in lumbar CSF and an isotonic solution. All samples were scanned simultaneously using the variable flip angle method with B1 correction, repeated six times on one 3T scanner, and once on a second 3T scanner. Difference in relaxivity between CSF and the isotonic solution were evaluated from the repeated measurements.

Results

There was a significant difference in relaxivity between CSF and the isotonic solution for both gadobutrol and gadoteric acid. The relaxivity for gadobutrol for the respective scanners was estimated to 3.02 ± 0.09 vs. 3.63 L mmol −1 s −1 in CSF and 2.35 ± 0.05 vs. 2.74 L mmol −1 s −1 in isotonic solution. For gadoteric acid, corresponding results were 2.47 ± 0.02 vs. 2.91 L mmol −1 s −1 in CSF and 2.37 ± 0.03 vs. 2.8 L mmol −1 s −1 in isotonic solution. Between the scanners, there was a high correlation ( R 2 0.998) but an 18% scaling difference in the T1 relaxation rates and corresponding relaxivities.

Conclusions

The relaxivity was higher in CSF than in the isotonic solution, particularly for gadobutrol. Systematic differences in relaxivity between scanners may potentially be corrected using a scaling factor derived from the T1 time of baseline CSF. For CSF studies using T1 mapping with a gadolinium‐based contrast agent, we recommend using a CSF‐specific relaxivity constant.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Organometallic CompoundsContrast MediaMagnetic Resonance ImagingCerebrospinal FluidCyclobutanesHeterocyclic CompoundsHumansPhantoms, ImagingReproducibility of Results

Stichwörter der Autor:innen

MRI contrast mediaT1 mappingcerebrospinal fluidphantomrelaxivity

Substanzen

Organometallic CompoundsgadobutrolContrast MediaCyclobutanesHeterocyclic Compounds

Autor:innen und Institutionen

  • Department of Diagnostics and Intervention, Biomedical Engineering and Radiation Physics, Umeå University, Umeå, Sweden.
  • Department of Diagnostics and Intervention, Diagnostic Radiology, Umeå University, Umeå, Sweden.
  • Department of Clinical Sciences, Neurosciences, Umeå University, Umeå, Sweden.
  • Umeå Center for Functional Brain Imaging (UFBI), Umeå University, Umeå, Sweden.
  • Department of Applied Physics and Electronics, Umeå University, Umeå, Sweden.

ORCID-Profile: Sofia Behndig, Anders Garpebring, Daniel Dahlgren Lindström

Förderung

  • Vetenskapsrådet
  • Stiftelsen för Strategisk Forskning

Zitieren

APA

Behndig, S., Garpebring, A., Dahlgren Lindström, D., Malm, J., Wåhlin, A., & Eklund, A. (2026). Relaxivity of Gadobutrol and Gadoteric Acid in Cerebrospinal Fluid at 3 T. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.70284

Vancouver

Behndig S, Garpebring A, Dahlgren Lindström D, Malm J, Wåhlin A, Eklund A. Relaxivity of Gadobutrol and Gadoteric Acid in Cerebrospinal Fluid at 3 T. Magnetic Resonance in Medicine. 2026. doi:10.1002/mrm.70284

BibTeX
@article{Behndig2026,
  author = {Behndig, Sofia and Garpebring, Anders and Dahlgren Lindström, Daniel and Malm, Jan and Wåhlin, Anders and Eklund, Anders},
  title = {Relaxivity of Gadobutrol and Gadoteric Acid in Cerebrospinal Fluid at 3 T},
  journal = {Magnetic Resonance in Medicine},
  year = {2026},
  doi = {10.1002/mrm.70284}
}
Kostenlos starten →