Simultaneous Multi‐Slice Acceleration for Free‐Breathing Motion Corrected Late Gadolinium Enhancement Imaging

Grzegorz Tomasz Kowalik iD, Karl P Kunze, Filippo Bosio, Peter Speier, Daniel Staeb iD, Radhouene Neji, Blanca Domenech-Ximenos, Andrew Tyler iD, Reza Razavi, Muhummad Sohaib Nazir, Amedeo Chiribiri, Sébastien Roujol iD

Magnetic Resonance in Medicine · 2026

DOI: 10.1002/mrm.70200 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2026. Verfasst von Grzegorz Tomasz Kowalik und 11 weiteren Autor:innen. Beteiligte Einrichtung: Faculty of Life Sciences and Medicine, School of Biomedical Engineering and Imaging Sciences, King's College London, London, UK. Hauptthemen laut MEDLINE: Gadolinium, Contrast Media, Heart, Magnetic Resonance Imaging. Die Arbeit nennt 38 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Biomedical Research Centre at Guy's and St Thomas' National Health Service (NHS) Foundation Trust, King's College London, British Heart Foundation (BHF).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

The objective of this study was to develop and evaluate an accelerated free‐breathing motion corrected and averaged late gadolinium enhancement MRI (FB PSIR‐MoCo LGE) protocols.

Methods

Simultaneous multi‐slice (SMS) bSSFP imaging with GC‐LOLA correction and a multiband factor of two was implemented into a FB PSIR‐MoCo LGE sequence with phase sensitive inversion recovery (PSIR) reconstruction. Two FB PSIR‐MoCo LGE protocols were acquired in 26 patients referred for clinical cardiac MRI: (1) SMS‐accelerated, and (2) non‐SMS reference. Myocardial sharpness, scar‐to‐blood CNR estimate (CNRe), scar‐to‐healthy myocardium CNRe, blood‐to‐healthy myocardium CNRe, and scar volume were assessed. Scar assessment was conducted in all LGE‐positive patients ( N = 10). Image quality (4‐point scale: 1— poor/non diagnostic , 4— excellent/diagnostic ) was assessed by consensus of two readers.

Results

SMS resulted in a twofold reduction of acquisition time (reference: 3.7 ± 0.9 min, proposed: 1.9 ± 0.6 min, p < 0.001). Although both approaches resulted in good to excellent image quality, the proposed approach resulted in slightly reduced image quality scores (reference: 3.64 ± 0.36 vs. proposed: 3.46 ± 0.36, p = 0.001) and modest (∼20%) reduction in CNRe. Myocardial sharpness measures showed no statistically significant differences between the two approaches (0.37 ± 0.06 vs. 0.37 ± 0.05 mm −1 , p = 0.84). Importantly, both approaches were in excellent agreement for scar volume assessment (bias = 0.48 ± 0.86 mL, R 2 = 0.996).

Conclusion

The proposed SMS FB PSIR‐MoCo LGE protocol reduced scan time by a factor of 2 when 16 slices were obtained, with minimal impact on image quality and CNRe, and no differences in myocardial sharpness. Importantly, it provided excellent agreement for myocardial scar assessment, in comparison to the reference FB PSIR‐MoCo LGE.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

GadoliniumContrast MediaHeartMagnetic Resonance ImagingImage Processing, Computer-AssistedHumansFemaleMiddle AgedMaleRespirationImage EnhancementImage Interpretation, Computer-AssistedMotionAdultAgedMyocardiumAlgorithmsReproducibility of Results

Stichwörter der Autor:innen

free‐breathinglate gadolinium enhancementmotion correctionsimultaneous multi‐slice

Substanzen

GadoliniumContrast Media

Autor:innen und Institutionen

  • Faculty of Life Sciences and Medicine, School of Biomedical Engineering and Imaging Sciences, King's College London, London, UK.
  • MR Research Collaborations, Siemens Healthcare Limited, Camberley, UK.
  • Cardiovascular Predevelopment, Siemens Healthcare GmbH, Erlangen, Germany.
  • MR Research Collaborations, Siemens Healthcare Limited, Melbourne, Australia.
  • Department of Radiology, Hospital Clínic de Barcelona, Barcelona, Spain.
  • Royal Brompton Hospital, Part of Guy's and St Thomas Hospital NHS Foundation Trust, London, UK.

ORCID-Profile: Grzegorz Tomasz Kowalik, Daniel Staeb, Andrew Tyler, Sébastien Roujol

Förderung

  • Biomedical Research Centre at Guy's and St Thomas' National Health Service (NHS) Foundation Trust
  • King's College London
  • British Heart Foundation (BHF)
  • National Institute for Health Research (NIHR)
  • Innovate UK
  • Foundation Trust
  • Engineering and Physical Sciences Research Council (EPSRC)
  • Wellcome Trust

Zitieren

APA

Kowalik, G. T., Kunze, K. P., Bosio, F., Speier, P., Staeb, D., Neji, R., Domenech-Ximenos, B., Tyler, A., Razavi, R., Nazir, M. S., Chiribiri, A., & Roujol, S. (2026). Simultaneous Multi‐Slice Acceleration for Free‐Breathing Motion Corrected Late Gadolinium Enhancement Imaging. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.70200

Vancouver

Kowalik GT, Kunze KP, Bosio F, Speier P, Staeb D, Neji R, et al. Simultaneous Multi‐Slice Acceleration for Free‐Breathing Motion Corrected Late Gadolinium Enhancement Imaging. Magnetic Resonance in Medicine. 2026. doi:10.1002/mrm.70200

BibTeX
@article{Kowalik2026,
  author = {Kowalik, Grzegorz Tomasz and Kunze, Karl P and Bosio, Filippo and Speier, Peter and Staeb, Daniel and Neji, Radhouene and Domenech-Ximenos, Blanca and Tyler, Andrew and Razavi, Reza and Nazir, Muhummad Sohaib and Chiribiri, Amedeo and Roujol, Sébastien},
  title = {Simultaneous Multi‐Slice Acceleration for Free‐Breathing Motion Corrected Late Gadolinium Enhancement Imaging},
  journal = {Magnetic Resonance in Medicine},
  year = {2026},
  doi = {10.1002/mrm.70200}
}
Kostenlos starten →