Temperature Mapping During MR ‐Guided Cryoablation Using a FLORET UTE Sequence
Magnetic Resonance in Medicine · 2026
DOI: 10.1002/mrm.70152 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2026. Verfasst von Moritz Gutt und 6 weiteren Autor:innen. Beteiligte Einrichtung: Institute of Diagnostic and Interventional Radiology, Hannover Medical School, Hannover, Germany. Hauptthemen laut MEDLINE: Cryosurgery, Liver, Magnetic Resonance Imaging, Thermometry. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 19 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Bundesministerium für Bildung und Forschung, Siemens Healthineers.
Abstract (Originalsprache)
Purpose
To implement a three‐dimensional FLORET‐UTE sequence for rapid MR thermometry during cryoablation and to validate voxel‐wise temperature maps against fiber‐optic probes in ex vivo bovine liver.
Methods
A FLORET‐UTE trajectory (TE 50 μs, 1.6 mm isotropic, 18 s/volume) was executed on a 3 T scanner while dry‐ice cryoablation was applied to liver phantoms. Four fiber‐optic sensors provided reference temperatures. Magnitude signal was normalized to baseline and fitted to a mono‐exponential temperature model in the first experiment; the unchanged model was applied prospectively to the second dataset. Images were reconstructed offline with ‐regularized non‐Cartesian SENSE. Accuracy below 0°C was assessed with root‐mean‐square error (RMSE) and Bland–Altman analysis.
Results
Calibration yielded a strong signal–temperature relationship (). Prospective validation produced RMSEs of 0.93°C–1.73°C (overall 1.33°C). Bland–Altman bias was °C with limits of agreement °C to 2.23°C. Serial temperature maps captured three‐dimensional ice‐ball dynamics at 18‐s temporal resolution—substantially faster than previously reported radial‐UTE and STIR‐UTE approaches.
Conclusion
FLORET‐UTE enables quantitative 3‐D MR thermometry of frozen tissue with < 20 s temporal resolution and < 2°C accuracy, overcoming the speed limitations of earlier UTE methods. The technique is a promising candidate for real‐time MR‐guided cryoablation and warrants in vivo evaluation.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Institute of Diagnostic and Interventional Radiology, Hannover Medical School, Hannover, Germany.
- Research Campus STIMULATE, Magdeburg, Germany.
ORCID-Profile: Moritz Gutt, Othmar Belker, Simon Schröer
Förderung
- Bundesministerium für Bildung und Forschung
- Siemens Healthineers
Zitieren
APA
Gutt, M., Belker, O., Scheller, J., Schröer, S., Wacker, F., Hensen, B., & Gutberlet, M. (2026). Temperature Mapping During MR ‐Guided Cryoablation Using a FLORET UTE Sequence. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.70152
Vancouver
Gutt M, Belker O, Scheller J, Schröer S, Wacker F, Hensen B, et al. Temperature Mapping During MR ‐Guided Cryoablation Using a FLORET UTE Sequence. Magnetic Resonance in Medicine. 2026. doi:10.1002/mrm.70152
BibTeX
@article{Gutt2026,
author = {Gutt, Moritz and Belker, Othmar and Scheller, Julienne and Schröer, Simon and Wacker, Frank and Hensen, Bennet and Gutberlet, Marcel},
title = {Temperature Mapping During MR ‐Guided Cryoablation Using a FLORET UTE Sequence},
journal = {Magnetic Resonance in Medicine},
year = {2026},
doi = {10.1002/mrm.70152}
}
EUVIMED