A unipolar head gradient for high‐field MRI without encoding ambiguity
Magnetic Resonance in Medicine · 2026
DOI: 10.1002/mrm.70098 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2026. Verfasst von Markus Weiger und 18 weiteren Autor:innen. Beteiligte Einrichtung: Institute for Biomedical Engineering, ETH Zurich and University of Zurich, Zurich, Switzerland. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Brain, Head. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 78 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
MRI gradients with a conventional, bipolar design generally face a trade‐off among performance, encoding ambiguity, and radiofrequency selectivity used to circumvent said ambiguity. This problem is particularly limiting in cutting‐edge brain imaging performed at field strengths ≥ 7 T and using high‐performance head gradients.
Methods
To address this issue, the present work proposes to fundamentally eliminate the encoding ambiguity in head gradients by using a unipolar z‐gradient design that takes advantage of the signal‐free range on one side of the imaging volume. This concept is demonstrated by implementation of a unipolar head gradient for operation at 7 T .
Results
Imaging in phantoms and in vivo demonstrates elimination of backfolding due to encoding ambiguity. At the same time, the unipolar design achieves efficiency on par with conventional bipolar design, resulting in high amplitude and slew‐rate performance.
Conclusion
The prospect of gradient systems based on a unipolar design holds promise for all advanced neuroimaging that demands high gradient performance. It will make the greatest difference at 7 T and beyond, where the absence of ambiguity removes a key concern and constraint in terms of radiofrequency behavior and instrumentation.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Institute for Biomedical Engineering, ETH Zurich and University of Zurich, Zurich, Switzerland.
- Uelzen, Germany.
- Institute for Energy and Process Engineering, ETH Zurich, Zurich, Switzerland.
- Philips AG, Zurich, Switzerland.
- Philips Healthcare, Best, The Netherlands.
- Futura Composites BV, Heerhugowaard, The Netherlands.
ORCID-Profile: Markus Weiger, Franciszek Hennel, Emily Louise Baadsvik, Oskar Björkqvist, Roger Luechinger, Eric Seth Michael, Klaas Paul Pruessmann
Zitieren
APA
Weiger, M., Overweg, J., Hennel, F., Baadsvik, E. L., Bianchi, S., Björkqvist, O., Luechinger, R., Metzger, J., Michael, E. S., Schmid, T., Singenberger, L., Sturzenegger, U., Oskam, E., Vissers, G., Koonen, J., Schuth, W., Koeleman, J., Borgo, M., & Pruessmann, K. P. (2026). A unipolar head gradient for high‐field MRI without encoding ambiguity. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.70098
Vancouver
Weiger M, Overweg J, Hennel F, Baadsvik EL, Bianchi S, Björkqvist O, et al. A unipolar head gradient for high‐field MRI without encoding ambiguity. Magnetic Resonance in Medicine. 2026. doi:10.1002/mrm.70098
BibTeX
@article{Weiger2026,
author = {Weiger, Markus and Overweg, Johan and Hennel, Franciszek and Baadsvik, Emily Louise and Bianchi, Samuel and Björkqvist, Oskar and Luechinger, Roger and Metzger, Jens and Michael, Eric Seth and Schmid, Thomas and Singenberger, Lauro and Sturzenegger, Urs and Oskam, Erik and Vissers, Gerrit and Koonen, Jos and Schuth, Wout and Koeleman, Jeroen and Borgo, Martino and Pruessmann, Klaas Paul},
title = {A unipolar head gradient for high‐field MRI without encoding ambiguity},
journal = {Magnetic Resonance in Medicine},
year = {2026},
doi = {10.1002/mrm.70098}
}
EUVIMED