Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2025. Verfasst von Tim Schmidt und 1 weiteren Autor:innen. Beteiligte Einrichtung: Laboratory for Social and Neural Systems Research (SNS Lab), University of Zurich, Zurich, Switzerland. Hauptthemen laut MEDLINE: Diffusion Magnetic Resonance Imaging, Quality Assurance, Health Care, Image Interpretation, Computer-Assisted, Artifacts. Die Arbeit nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
The aim of the presented work is to develop a diffusion instability measure (DIM) that can be used in quality assurance.
Methods
Single‐shot, spin‐echo, echo‐planar imaging HARDI diffusion data sets were collected on a spherical silicone oil phantom with 64 different diffusion directions on a 3T Philips Achieva and a 3T Siemens Cima.X scanner with similar acquisition protocols. A few data sets on the Philips Achieva included concurrent magnetic field monitoring. A correlation coefficient matrix among the diffusion directions for each data set was calculated, and subsequently, one minus its eigenratio was defined as the DIM.
Results
The DIM values ranged between about 5000 and 280 000 ppm across the data sets. The worst and best image quality—and hence the highest and lowest DIM values—were observed for b‐value = 4000 s/mm 2 on the Philips Achieva and for b‐value = 500 s/mm 2 on the Siemens Cima.X without concurrent field monitoring, respectively.
Conclusion
DIM is a sensitive tool for evaluating image quality in HARDI scans on a quantitative basis. It is simple to implement without the need for hardware or software modifications.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Laboratory for Social and Neural Systems Research (SNS Lab), University of Zurich, Zurich, Switzerland.
- Institute for Biomedical Engineering, ETH Zurich and University of Zurich, Zurich, Switzerland.
ORCID-Profile: Tim Schmidt, Zoltan Nagy
Zitieren
APA
Schmidt, T., & Nagy, Z. (2025). DIM : A diffusion instability measure for MRI quality assurance. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.70012
Vancouver
Schmidt T, Nagy Z. DIM : A diffusion instability measure for MRI quality assurance. Magnetic Resonance in Medicine. 2025. doi:10.1002/mrm.70012
BibTeX
@article{Schmidt2025,
author = {Schmidt, Tim and Nagy, Zoltan},
title = {DIM : A diffusion instability measure for MRI quality assurance},
journal = {Magnetic Resonance in Medicine},
year = {2025},
doi = {10.1002/mrm.70012}
}
EUVIMED