EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

3D Cartesian ultrashort double half‐echo imaging of the lung parenchyma for water density imaging

Richard B Thompson iD, Christopher Keen, Richard Coulden, Hefin Jones, Robert W Stobbe iD, Justin G Grenier

Magnetic Resonance in Medicine · 2026

DOI: 10.1002/mrm.30615 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2026. Verfasst von Richard B Thompson und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology and Diagnostic Imaging, University of Alberta, Edmonton, Alberta, Canada. Hauptthemen laut MEDLINE: Lung, Imaging, Three-Dimensional, Magnetic Resonance Imaging. Die Arbeit nennt 35 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Canadian Institutes for Health Research (CIHR), Canadian Institutes of Health Research.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Develop and illustrate a 3D double half‐echo Cartesian UTE method for spin‐density weighted imaging of the lung parenchyma and calculation of lung water density (LWD).

Methods

A 3D gradient‐echo pulse sequence was modified to acquire half‐echoes, to enable UTEs (TE/TR = 145 μs/1.2 ms), with an acquired resolution of 3.125 mm by 3.125 mm by 5 mm. Breath‐hold (12.9 s) and free‐breathing (94 s) acquisitions, using a center of k‐space navigator, were compared to a previously validated yarnball UTE sequence (1.5T/2.89T). Apparent SNR in the lung parenchyma was measured for all in‐vivo acquisitions. Illustrative clinical cases included heart failure and sarcoidosis with a comparison to CT images.

Results

Lung image quality and calculated LWD was similar for all compared methods at 1.5T and 2.89T for breath‐hold and free‐breathing acquisitions ( N = 10, p > 0.05), with no visible artifacts. The mean lung parenchyma SNR values were 18.4 ± 1.4, 21.8 ± 1.7 and 15.1 ± 1.0 for 1.5T free‐breathing, 2.89T free‐breathing and 2.89T breath‐hold, respectively, and 20.7 ± 1.1 for yarnball acquisitions (2.89T), with corresponding average LWD values of 26.7 ± 2.9%, 27.1 ± 2.5%, 27.1 ± 2.1% and 27.7 ± 2.7%. MRI LWD images and CT scans yielded similar image contrast and normalized signal intensity units. All Cartesian UTE images were reconstructed on the scanner without the requirement for gridding.

Conclusions

A double half‐echo Cartesian UTE pulse sequence provides water‐density weighted images of the lung parenchyma in a breath‐hold or short free‐breathing acquisition with sufficient signal to noise for quantification of LWD at 1.5T or 2.89T.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

LungImaging, Three-DimensionalMagnetic Resonance ImagingHumansMaleFemaleAdultTomography, X-Ray ComputedMiddle AgedHeart FailureAlgorithmsBreath HoldingSignal-To-Noise RatioWaterImage Processing, Computer-Assisted

Stichwörter der Autor:innen

UTElunglung water density (LWD)magnetic resonance imagingnon‐Cartesianpulmonary edema

Substanzen

Water

Autor:innen und Institutionen

  • Department of Radiology and Diagnostic Imaging, University of Alberta, Edmonton, Alberta, Canada.
  • Department of Biomedical Engineering, University of Alberta, Edmonton, Alberta, Canada.

ORCID-Profile: Richard B Thompson, Robert W Stobbe

Förderung

  • Canadian Institutes for Health Research (CIHR)
  • Canadian Institutes of Health Research

Zitieren

APA

Thompson, R. B., Keen, C., Coulden, R., Jones, H., Stobbe, R. W., & Grenier, J. G. (2026). 3D Cartesian ultrashort double half‐echo imaging of the lung parenchyma for water density imaging. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30615

Vancouver

Thompson RB, Keen C, Coulden R, Jones H, Stobbe RW, Grenier JG. 3D Cartesian ultrashort double half‐echo imaging of the lung parenchyma for water density imaging. Magnetic Resonance in Medicine. 2026. doi:10.1002/mrm.30615

BibTeX
@article{Thompson2026,
  author = {Thompson, Richard B and Keen, Christopher and Coulden, Richard and Jones, Hefin and Stobbe, Robert W and Grenier, Justin G},
  title = {3D Cartesian ultrashort double half‐echo imaging of the lung parenchyma for water density imaging},
  journal = {Magnetic Resonance in Medicine},
  year = {2026},
  doi = {10.1002/mrm.30615}
}
Kostenlos starten →