Validation of quantitative transport mapping (QTM) with an ex vivo perfused liver model
Magnetic Resonance in Medicine · 2025
DOI: 10.1002/mrm.30581 ↗Kurz gefasst
Validierungsstudie in der Zeitschrift Magnetic Resonance in Medicine, 2025. Verfasst von Dominick Romano und 12 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical Engineering, Cornell University, Ithaca, New York, USA. Hauptthemen laut MEDLINE: Liver, Magnetic Resonance Imaging, Image Processing, Computer-Assisted. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 47 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NIBIB NIH HHS, NIH HHS, NIDDK NIH HHS.
Abstract (Originalsprache)
Purpose
To evaluate the accuracy of a deep learning–based quantitative transport mapping method (QTMnet) for measuring total tissue perfusion.
Methods
QTMnet obtains tissue perfusion parametric maps from dynamic contrast‐enhanced MRI images by training on simulated data. This data uses synthetic arterial and venous vasculature geometries with flow based on constrained constructive optimization. Gadolinium contrast agent distribution is governed by the transport‐forward problem, allowing us to generate a synthetic concentration spacetime profile for a given flow, blood volume fraction, and boundary condition. Tissue flows determined by QTMnet were compared to those obtained with traditional perfusion quantification (Kety equation with Tofts generalization), which uses a global arterial input function. Their total flow accuracies were validated on explanted porcine livers that were connected to an MR compatible flow pump with specified total flow rate for dynamic contrast‐enhanced MRI experiments.
Results
The mean total flow error for QTMnet was −0.34% ± 16.21% with range [−24.79%, 23.96%], compared to −35.74% ± 36.30% [−77.28%, 29.87%] for the Kety method. QTMnet provides 72% lower mean absolute error than the Kety method (12.15% vs. 43.21%, a 3.6‐fold reduction).
Conclusion
The fluid mechanics‐based QTMnet accurately estimates total tissue flow in liver explants.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Biomedical Engineering, Cornell University, Ithaca, New York, USA.
- Department of Radiology, Weill Cornell Medical College, New York, New York, USA.
- Department of Electrical and Computer Engineering, Cornell University, Ithaca, New York, USA.
- Department of Surgery, Weill Cornell Medical College, New York, New York, USA.
- Department of Mechanical and Aerospace Engineering, Cornell University, Ithaca, New York, USA.
- Department of Radiology, Columbia University Vagelos College of Physicians & Surgeons, New York, New York, USA.
ORCID-Profile: Dominick Romano, Qihao Zhang, Alexandra G Roberts, Benjamin Weppner, Ilhami Kovanlikaya, Mert Şişman, Renjiu Hu, Thanh D Nguyen, Alexey V Dimov, Martin R Prince, Pascal Spincemaille, Yi Wang
Förderung
- NIBIB NIH HHS
- NIH HHS
- NIDDK NIH HHS
- NIH
Zitieren
APA
Romano, D., Zhang, Q., Roberts, A. G., Weppner, B., Shou, J., Kovanlikaya, I., Şişman, M., Hu, R., Nguyen, T. D., Dimov, A. V., Prince, M. R., Spincemaille, P., & Wang, Y. (2025). Validation of quantitative transport mapping (QTM) with an ex vivo perfused liver model. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30581
Vancouver
Romano D, Zhang Q, Roberts AG, Weppner B, Shou J, Kovanlikaya I, et al. Validation of quantitative transport mapping (QTM) with an ex vivo perfused liver model. Magnetic Resonance in Medicine. 2025. doi:10.1002/mrm.30581
BibTeX
@article{Romano2025,
author = {Romano, Dominick and Zhang, Qihao and Roberts, Alexandra G and Weppner, Benjamin and Shou, Jian and Kovanlikaya, Ilhami and Şişman, Mert and Hu, Renjiu and Nguyen, Thanh D and Dimov, Alexey V and Prince, Martin R and Spincemaille, Pascal and Wang, Yi},
title = {Validation of quantitative transport mapping (QTM) with an ex vivo perfused liver model},
journal = {Magnetic Resonance in Medicine},
year = {2025},
doi = {10.1002/mrm.30581}
}
EUVIMED