TE optimization for J‐difference editing of 2‐hydroxyglutarate at 3T
Magnetic Resonance in Medicine · 2025
DOI: 10.1002/mrm.30561 ↗At a glance
Journal article in Magnetic Resonance in Medicine, 2025. Written by Kimberly L Chan and 4 co-authors. Institution: Advanced Imaging Research Center, University of Texas Southwestern Medical Center, Dallas, Texas, USA. Major MEDLINE topics: Glutarates, Glioma, Brain Neoplasms, Magnetic Resonance Imaging. Cited 2 times so far (Crossref); 50 references. The full text is freely available (open access, CC BY 4.0). Funded by: Cancer Prevention and Research Institute of Texas (CPRIT).
Abstract (original language)
Purpose
To investigate the TE dependence of the edited 2‐hydroxyglutarate (2HG) signal, its separation from co‐edited glutamate plus glutamine (Glx), and fit accuracy in the presence of nuisance signals using a MEGA‐PRESS sequence.
Methods
Simulations were performed at TEs 70–160 ms to assess the signal intensity and 2HG‐Glx overlap as a function of TE. The effect of the 2HG‐Glx spectral overlap on the fit accuracy of 2HG was evaluated on simulated 2HG‐edited spectra with in vivo parameter variations. Data were acquired at TEs of 70 and 90 ms in 13 glioma patients to estimate the TE‐dependence of the 2HG and Glx signal intensity and at a TE of 120 ms in eight glioma patients to estimate the in vivo 2HG, Glx, and water T2 relaxation times.
Results
A TE of 90 ms was found to produce a maximal 2HG integral, which was 23% larger than that at a TE of 70 ms in vivo without a significant increase in 2HG‐Glx overlap. Lipid and residual water were 26% and 16% lower, respectively, at a TE of 90 ms versus 70 ms. Fit‐quality numbers were 49% lower at a TE of 90 ms versus 70 ms, indicating enhanced fits at a TE of 90 ms. The in vivo T2 relaxation times of 2HG, Glx, and water were 264, 177, and 110 ms, respectively.
Conclusion
A TE of 90 ms was best with a maximal 2HG signal, minimal 2HG‐Glx overlap, and minimal residual water and lipid contamination.
Subjects
MeSH terms (MEDLINE)
Author keywords
Substances
Authors and institutions
- Advanced Imaging Research Center, University of Texas Southwestern Medical Center, Dallas, Texas, USA.
- Department of Psychiatry, University of Texas Southwestern Medical Center, Dallas, Texas, USA.
- Department of Biomedical Engineering, University of Texas Dallas, Dallas, Texas, USA.
- Department of Internal Medicine, University of Texas Southwestern Medical Center, Dallas, Texas, USA.
- Department of Neurology, UT Southwestern Medical Center, Dallas, Texas, USA.
- Department of Neurological Surgery, University of Texas Southwestern Medical Center, Dallas, Texas, USA.
ORCID profiles: Kimberly L Chan
Funding
- Cancer Prevention and Research Institute of Texas (CPRIT)
Cite
APA
Chan, K. L., Hapani, R., Maher, E. A., Patel, T. R., & Henning, A. (2025). TE optimization for J‐difference editing of 2‐hydroxyglutarate at 3T. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30561
Vancouver
Chan KL, Hapani R, Maher EA, Patel TR, Henning A. TE optimization for J‐difference editing of 2‐hydroxyglutarate at 3T. Magnetic Resonance in Medicine. 2025. doi:10.1002/mrm.30561
BibTeX
@article{Chan2025,
author = {Chan, Kimberly L and Hapani, Rutul and Maher, Elizabeth A and Patel, Toral R and Henning, Anke},
title = {TE optimization for J‐difference editing of 2‐hydroxyglutarate at 3T},
journal = {Magnetic Resonance in Medicine},
year = {2025},
doi = {10.1002/mrm.30561}
}
EUVIMED