Referenceless 4D flow MRI using radial balanced SSFP at 0. 6 T
Magnetic Resonance in Medicine · 2025
DOI: 10.1002/mrm.30503 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2025. Verfasst von Charles McGrath und 6 weiteren Autor:innen. Beteiligte Einrichtung: Institute for Biomedical Engineering, University and ETH Zurich, Zurich, Switzerland. Hauptthemen laut MEDLINE: Imaging, Three-Dimensional, Magnetic Resonance Imaging, Image Interpretation, Computer-Assisted, Magnetic Resonance Angiography. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 57 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Schweizerischer Nationalfonds zur Förderung der Wissenschaftlichen Forschung.
Abstract (Originalsprache)
Purpose
To implement four‐dimensional‐flow MRI using phase‐contrast balanced steady‐state free precession (bSSFP) at 0.6 T using a free‐running three‐dimensional (3D) radial trajectory and referenceless background phase correction.
Methods
A free‐running, wobbling Archimedean spiral approach including bipolar velocity‐encoding gradients (3D PC‐bSSFP) was implemented on a 0.6T prototype scanner. Bipolar rewinder gradients were added to ensure first‐moment nulling per repetition time. Velocity encoding was performed using a three‐point encoding scheme (i.e., omitting a reference measurement). Advanced computer simulations were carried out to validate the approach. Image reconstruction was performed using a locally low‐rank approach. Results for anatomical visualization and flow quantification were reconstructed separately with different regularization factors. Background phase correction was achieved using phase estimation on time‐averaged reconstructions. In vivo data were acquired in 6 healthy subjects during free breathing. Additional two‐dimensional (2D) phase‐contrast spoiled gradient‐echo (2D PC‐GRE) breath‐hold data were obtained for reference to compare flow values in the ascending aorta, descending aorta, and pulmonary trunk.
Results
Velocity data acquired with 3D PC‐bSSFP compared well with 2D PC‐GRE (root mean square error = 3.96 cm/s), with minor underestimation of velocities (−0.52 cm/s). Cardiac phase‐dependent signal‐to‐noise ratios normalized for differences in scan time and resolution between 3D PC‐bSSFP and 2D PC‐GRE demonstrate relatively steady values for 3D PC‐bSSFP when compared to 2D PC‐bSSFP with some reduction during phases of high flow.
Conclusion
Free‐running, referenceless, four‐dimensional‐flow MRI using radial 3D PC‐bSSFP is feasible on a lower‐field 0.6T system, producing adequate flow quantification while yielding simultaneously reasonable cine images for concurrent flow and functional assessment of the heart and great vessels.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Institute for Biomedical Engineering, University and ETH Zurich, Zurich, Switzerland.
- Philips Healthcare, Best, The Netherlands.
- Philips Innovative Technologies, Hamburg, Germany.
ORCID-Profile: Charles McGrath, Pietro Dirix, Vincent Vousten, Jouke Smink, Ece Ercan, Peter Börnert, Sebastian Kozerke
Förderung
- Schweizerischer Nationalfonds zur Förderung der Wissenschaftlichen Forschung
Zitieren
APA
McGrath, C., Dirix, P., Vousten, V., Smink, J., Ercan, E., Börnert, P., & Kozerke, S. (2025). Referenceless 4D flow MRI using radial balanced SSFP at 0. 6 T. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30503
Vancouver
McGrath C, Dirix P, Vousten V, Smink J, Ercan E, Börnert P, et al. Referenceless 4D flow MRI using radial balanced SSFP at 0. 6 T. Magnetic Resonance in Medicine. 2025. doi:10.1002/mrm.30503
BibTeX
@article{McGrath2025,
author = {McGrath, Charles and Dirix, Pietro and Vousten, Vincent and Smink, Jouke and Ercan, Ece and Börnert, Peter and Kozerke, Sebastian},
title = {Referenceless 4D flow MRI using radial balanced SSFP at 0. 6 T},
journal = {Magnetic Resonance in Medicine},
year = {2025},
doi = {10.1002/mrm.30503}
}
EUVIMED