Denoising complex‐valued diffusion MR images using a two‐step, nonlocal principal component analysis approach
Magnetic Resonance in Medicine · 2025-06 · 93 (6) · 2473-2487
DOI: 10.1002/mrm.30502 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, online erschienen am 13.03.2025. Verfasst von Xinyu Ye und 6 weiteren Autor:innen. Beteiligte Einrichtung: Center for Biomedical Imaging Research, School of Biomedical Engineering, Tsinghua University, Beijing, China. Hauptthemen laut MEDLINE: Diffusion Magnetic Resonance Imaging, Image Enhancement. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 51 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NIH HHS, National Key Research and Development Program of China, NIBIB NIH HHS.
Abstract (Originalsprache)
Purpose
To propose a two‐step, nonlocal principal component analysis (PCA) method and demonstrate its utility for denoising complex diffusion MR images with a few diffusion directions.
Methods
A two‐step denoising pipeline was implemented to ensure accurate patch selection even with high noise levels and was coupled with data preprocessing for g‐factor normalization and phase stabilization before data denoising with a nonlocal PCA algorithm. At the heart of our proposed pipeline was the use of a data‐driven optimal shrinkage algorithm to manipulate the singular values in a way that would optimally estimate the noise‐free signal. Our approach's denoising performances were evaluated using simulation and in vivo human data experiments. The results were compared with those obtained with existing local PCA‐based methods.
Results
In both simulation and human data experiments, our approach substantially enhanced image quality relative to the noisy counterpart, yielding improved performances for estimation of relevant diffusion tensor imaging metrics. It also outperformed existing local PCA‐based methods in reducing noise while preserving anatomic details. It also led to improved whole‐brain tractography relative to the noisy counterpart.
Conclusion
The proposed denoising method has the utility for improving image quality for diffusion MRI with a few diffusion directions and is believed to benefit many applications, especially those aiming to achieve high‐quality parametric mapping using only a few image volumes.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Center for Biomedical Imaging Research, School of Biomedical Engineering, Tsinghua University, Beijing, China.
- Wellcome Center for Integrative Neuroimaging, FMRIB, Nuffield Department of Clinical Neurosciences, University of Oxford, Oxford, UK.
- Department of Radiology and Imaging Sciences, University of Utah, Salt Lake City, Utah, USA.
- Tiantan Neuroimaging Center of Excellence, Beijing Tiantan Hospital, Capital Medical University, Beijing, China.
- Center for Magnetic Resonance Research, Radiology, Medical School, University of Minnesota, Minneapolis, Minnesota, USA.
ORCID-Profile: Xinyu Ye, Xiaodong Ma, Hua Guo, Kamil Uğurbil, Xiaoping Wu
Förderung
- NIH HHS
- National Key Research and Development Program of China
- NIBIB NIH HHS
- NINDS NIH HHS
- National Institutes of Health
Zitieren
APA
Ye, X., Ma, X., Pan, Z., Zhang, Z., Guo, H., Uğurbil, K., & Wu, X. (2025). Denoising complex‐valued diffusion MR images using a two‐step, nonlocal principal component analysis approach. Magnetic Resonance in Medicine, 93(6), 2473-2487. https://doi.org/10.1002/mrm.30502
Vancouver
Ye X, Ma X, Pan Z, Zhang Z, Guo H, Uğurbil K, et al. Denoising complex‐valued diffusion MR images using a two‐step, nonlocal principal component analysis approach. Magnetic Resonance in Medicine. 2025;93(6):2473-2487. doi:10.1002/mrm.30502
BibTeX
@article{Ye2025,
author = {Ye, Xinyu and Ma, Xiaodong and Pan, Ziyi and Zhang, Zhe and Guo, Hua and Uğurbil, Kamil and Wu, Xiaoping},
title = {Denoising complex‐valued diffusion MR images using a two‐step, nonlocal principal component analysis approach},
journal = {Magnetic Resonance in Medicine},
year = {2025},
volume = {93},
number = {6},
pages = {2473--2487},
doi = {10.1002/mrm.30502}
}