EUVIMEDOpdag EUVIMED
Crossref

Computational study of the effects of orthopedic plates on gradient‐induced peripheral nerve stimulation under MRI using electromagnetic and neurophysiological modeling

Lijian Yang, Xiaolin Yang, Hui Ye, Norbert Kaula, Yuanhu Jin, Jianfeng Zheng iD, Wolfgang Kainz, Ji Chen iD

Magnetic Resonance in Medicine · 2025-06 · 93 (6) · 2583-2594

DOI: 10.1002/mrm.30470 ↗

At a glance

Journal article in Magnetic Resonance in Medicine, published online on 2025-02-18. Written by Lijian Yang and 7 co-authors. Institution: Department of Electrical and Computer Engineering, University of Houston, Houston, Texas, USA. Major MEDLINE topics: Magnetic Resonance Imaging, Peripheral Nerves. Cited 6 times so far (Crossref); 36 references. Funded by: Division of Industrial Innovation and Partnerships, Directorate for Technology, Innovation and Partnerships.

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Purpose

This study investigates the impact of orthopedic plates on peripheral nerve stimulation (PNS) during MRI, focusing on how the presence of implants affects PNS thresholds.

Methods

A combination of anatomical electromagnetic and neurophysiological modeling was utilized. Electromagnetic fields in human body models were simulated, both with and without orthopedic plates. Simulations were performed under x ‐, y ‐, and z ‐axis gradient coils, with plates implanted at two clinically relevant locations. Nerve responses were modeled using an established neurophysiological model.

Results

The presence of orthopedic implants significantly influenced nerve stimulation, leading to reductions in stimulation thresholds of up to 80%. Some of the reduced thresholds were close to the PNS limits outlined in International Electrotechnical Commission (IEC) 60601‐2‐33, suggesting a considerably reduced safety margin compared to cases without implants.

Conclusion

Orthopedic implants can substantially lower the activation thresholds of nearby nerves, with some thresholds approaching the PNS limits defined in IEC 60601‐2‐33 for MRI gradient field. This finding indicates a reduced safety margin for patients with implants, highlighting the need for more comprehensive safety assessments.

Subjects

MeSH terms (MEDLINE)

Magnetic Resonance ImagingPeripheral NervesHumansComputer SimulationElectromagnetic FieldsElectric StimulationModels, NeurologicalProstheses and Implants

Author keywords

MRI gradient coil switchingelectromagnetic field simulationneurodynamic modelperipheral nerve stimulation

Authors and institutions

  • Department of Electrical and Computer Engineering, University of Houston, Houston, Texas, USA.
  • Department of Biology, Loyola University Chicago, Chicago, Illinois, USA.
  • Ncor, LLC, Arvada, Colorado, USA.
  • High Performance Computing for MRI Safety, Jasper, Georgia, USA.

ORCID profiles: Jianfeng Zheng, Ji Chen

Funding

  • Division of Industrial Innovation and Partnerships
  • Directorate for Technology, Innovation and Partnerships

Cite

APA

Yang, L., Yang, X., Ye, H., Kaula, N., Jin, Y., Zheng, J., Kainz, W., & Chen, J. (2025). Computational study of the effects of orthopedic plates on gradient‐induced peripheral nerve stimulation under MRI using electromagnetic and neurophysiological modeling. Magnetic Resonance in Medicine, 93(6), 2583-2594. https://doi.org/10.1002/mrm.30470

Vancouver

Yang L, Yang X, Ye H, Kaula N, Jin Y, Zheng J, et al. Computational study of the effects of orthopedic plates on gradient‐induced peripheral nerve stimulation under MRI using electromagnetic and neurophysiological modeling. Magnetic Resonance in Medicine. 2025;93(6):2583-2594. doi:10.1002/mrm.30470

BibTeX
@article{Yang2025,
  author = {Yang, Lijian and Yang, Xiaolin and Ye, Hui and Kaula, Norbert and Jin, Yuanhu and Zheng, Jianfeng and Kainz, Wolfgang and Chen, Ji},
  title = {Computational study of the effects of orthopedic plates on gradient‐induced peripheral nerve stimulation under MRI using electromagnetic and neurophysiological modeling},
  journal = {Magnetic Resonance in Medicine},
  year = {2025},
  volume = {93},
  number = {6},
  pages = {2583--2594},
  doi = {10.1002/mrm.30470}
}
Start for free →