EUVIMEDOpdag EUVIMED
Open access · CC BY 4.0

Fast, automated, real‐time 3D passive balloon catheter tracking during MRI ‐guided cardiac catheterization using orthogonal projection imaging and real‐time image‐based catheter detection

Grzegorz T Kowalik iD, Eric Kerfoot, Karl Kunze, Radhouene Neji, Tracy Moon, Nina Mellor, Reza Razavi, Kuberan Pushparajah, Sébastien Roujol iD

Magnetic Resonance in Medicine · 2025-01 · 93 (1) · 311-320

DOI: 10.1002/mrm.30265 ↗

At a glance

Journal article in Magnetic Resonance in Medicine, 2024-09. Written by Grzegorz T Kowalik and 8 co-authors. Institution: School of Biomedical Engineering and Imaging Sciences, Faculty of Life Sciences and Medicine, King's College London, London, UK. Major MEDLINE topics: Imaging, Three-Dimensional, Cardiac Catheterization, Algorithms. Cited 3 times so far (Crossref); 27 references. The full text is freely available (open access, CC BY 4.0). Funded by: Engineering and Physical Sciences Research Council, Siemens, Wellcome EPSRC Centre for Medical Engineering.

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Purpose

MRI‐guidance of cardiac catheterization is currently performed using one or multiple 2D imaging planes, which may be suboptimal for catheter navigation, especially in patients with complex anatomies. The purpose of the work was to develop a robust real‐time 3D catheter tracking method and 3D visualization strategy for improved MRI‐guidance of cardiac catheterization procedures.

Methods

A fast 3D tracking technique was developed using continuous acquisition of two orthogonal 2D‐projection images. Each projection corresponds to a gradient echo stack of slices with only the central k‐space lines being collected for each slice. To enhance catheter contrast, a saturation pulse is added ahead of the projection pair. An offline image processing algorithm was developed to identify the 2D coordinates of the balloon in each projection image and to estimate its corresponding 3D coordinates. Post‐processing includes background signal suppression using an atlas of background 2D‐projection images. 3D visualization of the catheter and anatomy is proposed using three live sagittal, coronal, and axial (MPR) views and 3D rendering. The technique was tested in a subset of a catheterization step in three patients undergoing MRI‐guided cardiac catheterization using a passive balloon catheter.

Results

The extraction of the catheter balloon 3D coordinates was successful in all patients and for the majority of time‐points (accuracy >96%). This tracking method enabled a novel 3D visualization strategy for passive balloon catheter, providing enhanced anatomical context during catheter navigation.

Conclusion

The proposed tracking strategy shows promise for robust tracking of passive balloon catheter and may enable enhanced visualization during MRI‐guided cardiac catheterization.

Subjects

MeSH terms (MEDLINE)

Imaging, Three-DimensionalCardiac CatheterizationAlgorithmsHumansReproducibility of ResultsMagnetic Resonance Imaging, InterventionalComputer SystemsSensitivity and SpecificityMagnetic Resonance ImagingPattern Recognition, AutomatedImage Interpretation, Computer-AssistedImage Enhancement

Author keywords

MR‐guidancecardiac catheterizationpassive trackingreal time

Authors and institutions

  • School of Biomedical Engineering and Imaging Sciences, Faculty of Life Sciences and Medicine, King's College London, London, UK.
  • MR Research Collaborations, Siemens Healthcare Limited, Camberley, UK.
  • Guy's and St Thomas' NHS Foundation Trust, London, UK.

ORCID profiles: Grzegorz T Kowalik, Sébastien Roujol

Funding

  • Engineering and Physical Sciences Research Council
  • Siemens
  • Wellcome EPSRC Centre for Medical Engineering
  • National Institute for Health Research (NIHR) Biomedical Research Center based at Guy's
  • St. Thomas' National Health Service (NHS) Foundation Trust and King's College London
  • British Heart Foundation
  • National Institute for Health and Care Research

Cite

APA

Kowalik, G. T., Kerfoot, E., Kunze, K., Neji, R., Moon, T., Mellor, N., Razavi, R., Pushparajah, K., & Roujol, S. (2025). Fast, automated, real‐time 3D passive balloon catheter tracking during MRI ‐guided cardiac catheterization using orthogonal projection imaging and real‐time image‐based catheter detection. Magnetic Resonance in Medicine, 93(1), 311-320. https://doi.org/10.1002/mrm.30265

Vancouver

Kowalik GT, Kerfoot E, Kunze K, Neji R, Moon T, Mellor N, et al. Fast, automated, real‐time 3D passive balloon catheter tracking during MRI ‐guided cardiac catheterization using orthogonal projection imaging and real‐time image‐based catheter detection. Magnetic Resonance in Medicine. 2025;93(1):311-320. doi:10.1002/mrm.30265

BibTeX
@article{Kowalik2025,
  author = {Kowalik, Grzegorz T and Kerfoot, Eric and Kunze, Karl and Neji, Radhouene and Moon, Tracy and Mellor, Nina and Razavi, Reza and Pushparajah, Kuberan and Roujol, Sébastien},
  title = {Fast, automated, real‐time 3D passive balloon catheter tracking during MRI ‐guided cardiac catheterization using orthogonal projection imaging and real‐time image‐based catheter detection},
  journal = {Magnetic Resonance in Medicine},
  year = {2025},
  volume = {93},
  number = {1},
  pages = {311--320},
  doi = {10.1002/mrm.30265}
}
Start for free →