On the impact of B0 shimming algorithms on single‐voxel MR spectroscopy
Magnetic Resonance in Medicine · 2025
DOI: 10.1002/mrm.30257 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2025. Verfasst von Behrouz Vejdani Afkham und 1 weiteren Autor:innen. Beteiligte Einrichtung: NeuroPoly Lab, Institute of Biomedical Engineering, Polytechnique Montréal, Montréal, Quebec, Canada. Hauptthemen laut MEDLINE: Algorithms, Magnetic Resonance Spectroscopy. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 33 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: Natural Sciences and Engineering Research Council of Canada, Institut TransMedTech, Polytechnique Montréal.
Abstract (Originalsprache)
Purpose
To assess the impact of different B0 shimming algorithms on MRS.
Methods
B0 field maps and single‐voxel MR spectroscopy were acquired in the prefrontal cortex of five volunteers at 3 T using five different B0 shimming approaches. B0 shimming was achieved using Siemens' proprietary shim algorithm, in addition to the Pseudo‐Inverse (PI), Quadratic Programming (QuadProg), Least Squares (LSq), and Gradient optimization (Grad) algorithms. The standard deviation of the shimmed B0 field, as well as the SNR and FWHM of the measured metabolites, was used to evaluate the performance of each B0 shimming algorithm.
Results
Compared to Siemens's shim, significant reductions ( p < 0.01) in the standard deviation of the B0 field distribution within the MRS voxel were observed for the PI, QuadProg, and Grad algorithms (3.8 Hz, 7.3 Hz, and 3.9 Hz respectively, compared to 11.5 Hz for Siemens), but not for the LSq (12.9 Hz) algorithm. Moreover, significantly increased SNR and reduced FWHM for the N ‐acetylaspartate metabolite were consistent with the improvement in B0 homogeneity for the aforementioned shimming algorithms.
Conclusion
Here, we demonstrate that the choice of B0 shimming algorithm can have a significant impact on the quality of MR spectra and that significant improvements in spectrum quality could be achieved by using alternatives to the default vendor approach.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- NeuroPoly Lab, Institute of Biomedical Engineering, Polytechnique Montréal, Montréal, Quebec, Canada.
- Centre de recherche du CHU Sainte-Justine, Montréal, Quebec, Canada.
ORCID-Profile: Behrouz Vejdani Afkham, Eva Alonso-Ortiz
Förderung
- Natural Sciences and Engineering Research Council of Canada
- Institut TransMedTech
- Polytechnique Montréal
Zitieren
APA
Vejdani Afkham, B., & Alonso-Ortiz, E. (2025). On the impact of B0 shimming algorithms on single‐voxel MR spectroscopy. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30257
Vancouver
Vejdani Afkham B, Alonso-Ortiz E. On the impact of B0 shimming algorithms on single‐voxel MR spectroscopy. Magnetic Resonance in Medicine. 2025. doi:10.1002/mrm.30257
BibTeX
@article{VejdaniAfkham2025,
author = {Vejdani Afkham, Behrouz and Alonso-Ortiz, Eva},
title = {On the impact of B0 shimming algorithms on single‐voxel MR spectroscopy},
journal = {Magnetic Resonance in Medicine},
year = {2025},
doi = {10.1002/mrm.30257}
}
EUVIMED