3D MR fingerprinting for dynamic contrast‐enhanced imaging of whole mouse brain

Yuran Zhu iD, Guanhua Wang iD, Yuning Gu, Walter Zhao iD, Jiahao Lu, Junqing Zhu, Christina J MacAskill iD, Andrew Dupuis iD, Mark A Griswold iD, Dan Ma iD, Chris A Flask iD, Xin Yu iD

Magnetic Resonance in Medicine · 2025

DOI: 10.1002/mrm.30253 ↗

At a glance

Journal article in Magnetic Resonance in Medicine, 2025. Written by Yuran Zhu and 11 co-authors. Institution: Department of Biomedical Engineering, Case Western Reserve University, Cleveland, Ohio, USA. Major MEDLINE topics: Contrast Media, Imaging, Three-Dimensional, Magnetic Resonance Imaging, Brain. Cited 3 times so far (Crossref); 56 references. The full text is freely available (open access, CC BY-NC-ND 4.0). Funded by: NIGMS NIH HHS, American Heart Association, NCATS NIH HHS.

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Purpose

Quantitative MRI enables direct quantification of contrast agent concentrations in contrast‐enhanced scans. However, the lengthy scan times required by conventional methods are inadequate for tracking contrast agent transport dynamically in mouse brain. We developed a 3D MR fingerprinting (MRF) method for simultaneous T 1 and T 2 mapping across the whole mouse brain with 4.3‐min temporal resolution.

Method

We designed a 3D MRF sequence with variable acquisition segment lengths and magnetization preparations on a 9.4T preclinical MRI scanner. Model‐based reconstruction approaches were employed to improve the accuracy and speed of MRF acquisition. The method's accuracy for T 1 and T 2 measurements was validated in vitro, while its repeatability of T 1 and T 2 measurements was evaluated in vivo ( n = 3). The utility of the 3D MRF sequence for dynamic tracking of intracisternally infused gadolinium‐diethylenetriamine pentaacetic acid (Gd‐DTPA) in the whole mouse brain was demonstrated ( n = 5).

Results

Phantom studies confirmed accurate T 1 and T 2 measurements by 3D MRF with an undersampling factor of up to 48. Dynamic contrast‐enhanced MRF scans achieved a spatial resolution of 192 × 192 × 500 μm 3 and a temporal resolution of 4.3 min, allowing for the analysis and comparison of dynamic changes in concentration and transport kinetics of intracisternally infused Gd‐DTPA across brain regions. The sequence also enabled highly repeatable, high‐resolution T 1 and T 2 mapping of the whole mouse brain (192 × 192 × 250 μm 3 ) in 30 min.

Conclusion

We present the first dynamic and multi‐parametric approach for quantitatively tracking contrast agent transport in the mouse brain using 3D MRF.

Subjects

MeSH terms (MEDLINE)

Contrast MediaImaging, Three-DimensionalMagnetic Resonance ImagingBrainGadolinium DTPAPhantoms, ImagingAnimalsMiceReproducibility of ResultsAlgorithmsMice, Inbred C57BLImage EnhancementImage Processing, Computer-Assisted

Author keywords

3DCSFMR fingerprintingT1T2dynamic contrast‐enhanced

Substances

Contrast MediaGadolinium DTPA

Authors and institutions

  • Department of Biomedical Engineering, Case Western Reserve University, Cleveland, Ohio, USA.
  • Department of Biomedical Engineering, University of Michigan, Ann Arbor, Michigan, USA.
  • Department of Radiology, Case Western Reserve University, Cleveland, Ohio, USA.
  • Department of Pediatrics, Case Western Reserve University, Cleveland, Ohio, USA.
  • Department of Physiology and Biophysics, Case Western Reserve University, Cleveland, Ohio, USA.

ORCID profiles: Yuran Zhu, Guanhua Wang, Walter Zhao, Christina J MacAskill, Andrew Dupuis, Mark A Griswold, Dan Ma, Chris A Flask, Xin Yu

Funding

  • NIGMS NIH HHS
  • American Heart Association
  • NCATS NIH HHS
  • NINDS NIH HHS
  • NIBIB NIH HHS

Cite

APA

Zhu, Y., Wang, G., Gu, Y., Zhao, W., Lu, J., Zhu, J., MacAskill, C. J., Dupuis, A., Griswold, M. A., Ma, D., Flask, C. A., & Yu, X. (2025). 3D MR fingerprinting for dynamic contrast‐enhanced imaging of whole mouse brain. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30253

Vancouver

Zhu Y, Wang G, Gu Y, Zhao W, Lu J, Zhu J, et al. 3D MR fingerprinting for dynamic contrast‐enhanced imaging of whole mouse brain. Magnetic Resonance in Medicine. 2025. doi:10.1002/mrm.30253