Hybrid algorithms for SAR matrix compression and the impact of post‐processing on SAR calculation complexity
Magnetic Resonance in Medicine · 2024
DOI: 10.1002/mrm.30235 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2024. Verfasst von Stephan Orzada und 2 weiteren Autor:innen. Beteiligte Einrichtung: Medical Physics in Radiology, German Cancer Research Center (DKFZ), Heidelberg, Germany. Hauptthemen laut MEDLINE: Algorithms, Magnetic Resonance Imaging, Data Compression. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 29 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: HORIZON EUROPE Framework Programme.
Abstract (Originalsprache)
Purpose
This study proposes faster virtual observation point (VOP) compression as well as post‐processing algorithms for specific absorption rate (SAR) matrix compression. Furthermore, it shows the relation between the number of channels and the computational burden for VOP‐based SAR calculation.
Methods
The proposed new algorithms combine the respective benefits of two different criteria for determining upper boundedness of SAR matrices by the VOPs. Comparisons of the old and new algorithms are performed for head coil arrays with various channel counts. The new post‐processing algorithm is used to post‐process the VOP sets of nine arrays, and the number of VOPs for a fixed median relative overestimation is compared.
Results
The new algorithms are faster than the old algorithms by a factor of two to more than 10. The compression efficiency (number of VOPs relative to initial number of SAR matrices) is identical. For a fixed median relative overestimation, the number of VOPs increases logarithmically with the number of RF coil channels when post‐processing is applied.
Conclusion
The new algorithms are much faster than previous algorithms. Post‐processing is very beneficial for online SAR supervision of MRI systems with high channel counts, since for a given number of VOPs the relative SAR overestimation can be lowered.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Medical Physics in Radiology, German Cancer Research Center (DKFZ), Heidelberg, Germany.
- Erwin L. Hahn Institute for MRI, University Duisburg-Essen, Essen, Germany.
- Faculty of Physics and Astronomy, Heidelberg University, Heidelberg, Germany.
- Faculty of Medicine, Heidelberg University, Heidelberg, Germany.
ORCID-Profile: Stephan Orzada, Thomas M Fiedler
Förderung
- HORIZON EUROPE Framework Programme
Zitieren
APA
Orzada, S., Fiedler, T. M., & Ladd, M. E. (2024). Hybrid algorithms for SAR matrix compression and the impact of post‐processing on SAR calculation complexity. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30235
Vancouver
Orzada S, Fiedler TM, Ladd ME. Hybrid algorithms for SAR matrix compression and the impact of post‐processing on SAR calculation complexity. Magnetic Resonance in Medicine. 2024. doi:10.1002/mrm.30235
BibTeX
@article{Orzada2024,
author = {Orzada, Stephan and Fiedler, Thomas M and Ladd, Mark E},
title = {Hybrid algorithms for SAR matrix compression and the impact of post‐processing on SAR calculation complexity},
journal = {Magnetic Resonance in Medicine},
year = {2024},
doi = {10.1002/mrm.30235}
}
EUVIMED