Comprehensive assessment of nonuniform image quality: Application to imaging near metal

Alexander R Toews iD, Philip K Lee iD, Krishna S Nayak iD, Brian A Hargreaves

Magnetic Resonance in Medicine · 2024

DOI: 10.1002/mrm.30222 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2024. Verfasst von Alexander R Toews und 3 weiteren Autor:innen. Beteiligte Einrichtung: Radiology, Stanford University, Stanford, California, USA. Hauptthemen laut MEDLINE: Phantoms, Imaging, Metals, Magnetic Resonance Imaging, Artifacts. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 41 Literaturangaben. Gefördert von: Natural Sciences and Engineering Research Council of Canada, NIBIB NIH HHS, NIAMS NIH HHS.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Comprehensive assessment of image quality requires accounting for spatial variations in (i) intensity artifact, (ii) geometric distortion, (iii) signal‐to‐noise ratio (SNR), and (iv) spatial resolution, among other factors. This work presents an ensemble of methods to meet this need, from phantom design to image analysis, and applies it to the scenario of imaging near metal.

Methods

A modular phantom design employing a gyroid lattice is developed to enable the co‐registered volumetric quantitation of image quality near a metallic hip implant. A method for measuring spatial resolution by means of local point spread function (PSF) estimation is presented and the relative fitness of gyroid and cubic lattices is examined. Intensity artifact, geometric distortion, and SNR maps are also computed. Results are demonstrated with 2D‐FSE and MAVRIC‐SL scan protocols on a 3T MRI scanner.

Results

The spatial resolution method demonstrates a worst‐case error of 0.17 pixels for measuring in‐plane blurring up to 3 pixels (full width at half maximum). The gyroid outperforms a cubic lattice design for the local PSF estimation task. The phantom supports four configurations toggling the presence/absence of both metal and structure with good spatial correspondence for co‐registered analysis of the four quality factors. The marginal scan time to evaluate one scan protocol amounts to five repetitions. The phantom design can be fabricated in 2 days at negligible material cost.

Conclusion

The phantom and associated analysis methods can elucidate complex image quality trade‐offs involving intensity artifact, geometric distortion, SNR, and spatial resolution. The ensemble of methods is suitable for benchmarking imaging performance near metal.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Phantoms, ImagingMetalsMagnetic Resonance ImagingArtifactsAlgorithmsHumansSignal-To-Noise RatioReproducibility of ResultsImage EnhancementSensitivity and SpecificityImage Interpretation, Computer-AssistedImage Processing, Computer-Assisted

Stichwörter der Autor:innen

image qualitymetalphantompoint spread functionspatial resolutionsusceptibility artifact

Substanzen

Metals

Autor:innen und Institutionen

  • Radiology, Stanford University, Stanford, California, USA.
  • Electrical Engineering, Stanford University, Stanford, California, USA.
  • Ming Hsieh Department of Electrical and Computer Engineering, University of Southern California, Los Angeles, California, USA.
  • Bioengineering, Stanford University, Stanford, California, USA.

ORCID-Profile: Alexander R Toews, Philip K Lee, Krishna S Nayak

Förderung

  • Natural Sciences and Engineering Research Council of Canada
  • NIBIB NIH HHS
  • NIAMS NIH HHS
  • National Institutes of Health

Zitieren

APA

Toews, A. R., Lee, P. K., Nayak, K. S., & Hargreaves, B. A. (2024). Comprehensive assessment of nonuniform image quality: Application to imaging near metal. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30222

Vancouver

Toews AR, Lee PK, Nayak KS, Hargreaves BA. Comprehensive assessment of nonuniform image quality: Application to imaging near metal. Magnetic Resonance in Medicine. 2024. doi:10.1002/mrm.30222

BibTeX
@article{Toews2024,
  author = {Toews, Alexander R and Lee, Philip K and Nayak, Krishna S and Hargreaves, Brian A},
  title = {Comprehensive assessment of nonuniform image quality: Application to imaging near metal},
  journal = {Magnetic Resonance in Medicine},
  year = {2024},
  doi = {10.1002/mrm.30222}
}
Kostenlos starten →