Efficient pulse sequence design framework for high‐dimensional MR fingerprinting scans using systematic error index
Magnetic Resonance in Medicine · 2024
DOI: 10.1002/mrm.30155 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2024. Verfasst von Siyuan Hu und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical Engineering, Case Western Reserve University, Cleveland, Ohio, USA. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Algorithms, Brain, Computer Simulation. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 46 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: NIGMS NIH HHS, NICHD NIH HHS, NCI NIH HHS.
Abstract (Originalsprache)
Purpose
For effective optimization of MR fingerprinting (MRF) pulse sequences, estimating and minimizing errors from actual scan conditions are crucial. Although virtual‐scan simulations offer an approximation to these errors, their computational demands become expensive for high‐dimensional MRF frameworks, where interactions between more than two tissue properties are considered. This complexity makes sequence optimization impractical. We introduce a new mathematical model, the systematic error index (SEI), to address the scalability challenges for high‐dimensional MRF sequence design.
Methods
By eliminating the need to perform dictionary matching, the SEI model approximates quantification errors with low computational costs. The SEI model was validated in comparison with virtual‐scan simulations. The SEI model was further applied to optimize three high‐dimensional MRF sequences that quantify two to four tissue properties. The optimized scans were examined in simulations and healthy subjects.
Results
The proposed SEI model closely approximated the virtual‐scan simulation outcomes while achieving hundred‐ to thousand‐times acceleration in the computational speed. In both simulation and in vivo experiments, the optimized MRF sequences yield higher measurement accuracy with fewer undersampling artifacts at shorter scan times than the heuristically designed sequences.
Conclusion
We developed an efficient method for estimating real‐world errors in MRF scans with high computational efficiency. Our results illustrate that the SEI model could approximate errors both qualitatively and quantitatively. We also proved the practicality of the SEI model of optimizing sequences for high‐dimensional MRF frameworks with manageable computational power. The optimized high‐dimensional MRF scans exhibited enhanced robustness against undersampling and system imperfections with faster scan times.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Biomedical Engineering, Case Western Reserve University, Cleveland, Ohio, USA.
- Department of Radiology, Case Western Reserve University, Cleveland, Ohio, USA.
- Department of Mathematics, Applied Mathematics, and Statistics, Case Western Reserve University, Cleveland, Ohio, USA.
ORCID-Profile: Siyuan Hu, Zhilang Qiu, Walter Zhao, Rasim Boyacioglu, Debra McGivney, Dan Ma
Förderung
- NIGMS NIH HHS
- NICHD NIH HHS
- NCI NIH HHS
- NIBIB NIH HHS
- NINDS NIH HHS
- National Institute of Neurological Disorders and Stroke
- National Institute of Biomedical Imaging and Bioengineering
- National Cancer Institute
Zitieren
APA
Hu, S., Qiu, Z., Adams, R. J., Zhao, W., Boyacioglu, R., Calvetti, D., McGivney, D., & Ma, D. (2024). Efficient pulse sequence design framework for high‐dimensional MR fingerprinting scans using systematic error index. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30155
Vancouver
Hu S, Qiu Z, Adams RJ, Zhao W, Boyacioglu R, Calvetti D, et al. Efficient pulse sequence design framework for high‐dimensional MR fingerprinting scans using systematic error index. Magnetic Resonance in Medicine. 2024. doi:10.1002/mrm.30155
BibTeX
@article{Hu2024,
author = {Hu, Siyuan and Qiu, Zhilang and Adams, Richard James and Zhao, Walter and Boyacioglu, Rasim and Calvetti, Daniela and McGivney, Debra and Ma, Dan},
title = {Efficient pulse sequence design framework for high‐dimensional MR fingerprinting scans using systematic error index},
journal = {Magnetic Resonance in Medicine},
year = {2024},